Simple working mode
If the current/pressure is smaller than preset threshold, the indicator on the display and the output
switch of related comparator opens contact on output connector and will stay unchanged until
current/pressure value does not go over threshold value.
If the current/pressure is bigger than preset threshold, the state of the indicator on the display and
the output switch of related comparator closes contact on output connector and will stay
unchanged until current/pressure value exceeds threshold value.
Window working mode
In this working mode we have two thresholds: min threshold and max threshold.
If the current/pressure is between preset min threshold and max threshold values, the state of the
indicator and output switch of related comparator on output connector are closed contacts and will
stay unchanged until current/pressure value exceeds max threshold value or until current/pressure
value goes under preset min threshold value.
The hysteresis switching on / off is given by adjusted values min threshold and max threshold.
To change min and max thresholds see SETTINGS chapter.
The pin layout of the output connector is described in the Sec.8.1.
4.4.

Operating IP in case of main line interruption

In case of interruption of line voltage, the SIP POWER restarts in the conditions indicated in
section 3.1.
For security reasons it is not possible to activate automatic restarts.
In case of line interruption any alarm indication is cleared.
USER MANUAL - SIP POWER ION PUMP CONTROLLER
M.HIST.0109.23 Rev.1
Page 17 of 62