SAES SIP POWER User Manual page 19

Ion pump controller
Table of Contents

Advertisement

Second character: setting and status of SW1 switch
In Stop status the displayed character indicates the selected setting.
A white space means SW1 is OFF, while "S" means SW1 is ON.
In Active status the symbol indicates the current status of the switch (as set on the connector
IN/OUT INTEFACE of back panel, see wiring diagram for details):
= open switch
and
Third character: setting and status of SW2 switch
In Stop status the displayed character indicates the selected setting.
A white space means SW2 is OFF, while "S" means that SW2 is ON and set to Simple working
mode and "W" means that SW2 is on and set to Window working mode.
In Active status the symbol indicates the current status of the switch (as set on the connector
IN/OUT INTEFACE of back panel, see wiring diagram for details):
= open switch
and
Fourth character: setting and status of SW3 switch
In Stop status the displayed character indicates the selected setting.
A white space means SW3 is OFF, while "S" means that SW3 is ON and set to Simple working
mode and "W" means that SW3 is on and set to Window working mode.
In Active status the symbol indicates the current status of the switch (as set on the connector
IN/OUT INTEFACE of back panel, see wiring diagram for details):
= open switch
and
Supply time
This field refers to the IP working time since the high voltage supply starts. For short times (less
than 1 hour) the value is shown in minutes ("m"), for longer times in hours ("h").
Current value of electrical variables
On the second row the operating values are shown.
The current flowing through the ion pump is shown on the left side of the display. The pump
voltage, output power and pump current can be alternatively displayed by pushing the pushbutton
A / V / W.
The pump voltage is shown in V (Volt), current measurement automatically sets to nA - uA - mA
-9
(nano (10
) - micro (10
mW or W (milli Watt (10
Note: the voltage here reported is the actual value measured on the output in real time, without any
average, and without any processing. So it is possible that it changes slightly instantly.
M.HIST.0109.23 Rev.1
= closed switch
= closed switch
= closed switch
-6
-3
) - milli (10
) - Ampère) with a resolution of about 10nA, power range is in
-3
W) or Watt)
USER MANUAL - SIP POWER ION PUMP CONTROLLER
Page 19 of 62