Urostym™
Service Manual
LABORIE UROSTYM Service Manual UST-SM01-V13.00

Advertisement

loading
Need help?

Need help?

Do you have a question about the Urostym and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for LABORIE Urostym

  • Page 1 Urostym™ Service Manual LABORIE UROSTYM Service Manual UST-SM01-V13.00...
  • Page 2 Tel.: +31.53.4803700 Fax: +31.53.4803701 TRADEMARKS Urostym is a trademark of LABORIE MEDICAL TECHNOLOGIES CANADA ULC. Other names may be trademarks of their respective owners. COPYRIGHT All rights reserved. No part of this publication may be reproduced in any form whatsoever without the prior written permission of Laborie Medical Technologies Canada ULC.
  • Page 3: Table Of Contents

    3.3.2 Recommended adhesion positioning for printer: ............... 12 User Replaceable Fuse ........................13 CALIBRATION ............................14 Calibrate Urocap Using the Urostym CheckDB Utility ..............14 Calibration tips ..........................15 Calibrating EMG/Manomety ....................... 9 CLEANING AND PREVENTATIVE MAINTENANCE ................10 Urostym Unit and Accessories ......................10 Urocap Uroflowmeter ........................
  • Page 4 SCHEMATIC/PCB LABOUT ......................... 22 END-USER LICENSE AGREEMENT ....................23 LABORIE UROSTYM Service Manual UST-SM01...
  • Page 5: Introduction

    1 INTRODUCTION LABORIE UROSTYM Service Manual UST-SM01...
  • Page 6: Technical Specifications

    Temperature: +10°C to +40°C Humidity: 30% to 75% relative humidity Pressure: 70 kPa to 106 kPa Temperature: -29°C to +60°C Humidity: uncontrolled to 85% relative humidity 2.1.1 Power supply 2.1.2 Pressure Transducer 2.1.3 Line Isolation Transformer LABORIE UROSTYM Service Manual UST-SM01...
  • Page 7: Channels

    Manometry 0-250 cmH2O (0- ±10% FS 184mmHq) Output Channels: CAUTION: The values listed in the table below are the maximum values that the UROSTYM unit will supply. Laborie Medical Technologies strongly recommends staying within the specified ranges. Channel Range Tolerance ±...
  • Page 8: Power Supply

    EN ISO 15223-1 EN61000-3-2 • • EN ISO 14155 • EN 61000-3-3 • EN ISO 10993-1 • EN 61000-4-2:2009 Please refer to the EC Declaration of Conformity for a list of additional product related standards. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 9: Identification Of Components

    3 IDENTIFICATION OF COMPONENTS EXTERNAL COMPONENTS LABORIE UROSTYM Service Manual UST-SM01...
  • Page 10: Internal Components

    All internal components are clearly identified and labeled below. Reference Number Description Fan Power Cables Analog Power Cables RS-232 Communications Cable UROSTYM Enclosure Digital Board Reset Button Cables Power Entry Module Power Board Ground Stud Terminal Analog Board LED Board LABORIE UROSTYM Service Manual UST-SM01...
  • Page 11: Cart Setup

    5. Ensure the laptop is secured to the Urostym® by attempting to slide the laptop along the Urostym®. If the laptop slides remove the laptop from the Urostym® and remove the re- closable fasteners applied.
  • Page 12: Recommended Adhesion Positioning For Printer

    Velcro tape from the printer and cart shelf, clean the areas with alcohol wipes and retry the procedure from step 2. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 13: User Replaceable Fuse

    If the replacement fuse immediately burns out, move the power switch to the ‘off’ position and remove the power cord from the unit. Contact your LABORIE Service representative for further instruction. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 14: Calibration

    4 CALIBRATION CALIBRATE UROCAP USING THE UROSTYM CHECKDB UTILITY If the UDS120 software is not installed on the computer, the Urocap transducer can be calibrated through the Urostym CheckDB utility. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 15: Calibration Tips

    To verify your calibration, choose a Uroflow protocol, set zeroes, and pour using the same pouring procedure as the calibration. The unit should provide you with the same volume reading as during calibration. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 16: Calibrating Emg/Manomety

    CALIBRATING EMG/MANOMETY To calibrate channels 1,2, and 3: 1. Click Options > EMG/Manomentry Calibration. • • • • • LABORIE UROSTYM Service Manual UST-SM01...
  • Page 17: Cleaning And Preventative Maintenance

    5 CLEANING AND PREVENTATIVE MAINTENANCE UROSTYM UNIT AND ACCESSORIES The Urostym unit is not immersible. It should be wiped down with a clean cloth dampened with acleaning solution as per your facility’s cleaning instructions.  • • • UROCAP UROFLOWMETER The Urocap may becoe dirty due to urine contamination and cleaning will be required.
  • Page 18: Probes And Surface Electrode Patches

    DATA BACKUP Check the space available on the computer's hard disk at least every 6 months. Back up data from the UROSTYM MDB file located in the Urostym folder on the computer (C:\Program Files\Laborie Medical Technologies\Urostym\) on external media like CDs, DVDs or USB flash drives.
  • Page 19: Troubleshooting Guide

    6 TROUBLESHOOTING GUIDE In Stage (2) of the “Recommendation” section on the following pages, the UROSTYM unit should be opened. Laborie Medical Technologies strongly recommends that only qualified Laborie trained technicians service the unit at this stage due to the high voltage present inside the unit. The tool to be used in this stage is a multimeter.
  • Page 20 DB9 connector to the digital between the PC and the • board. Check the “UROSTYM.ini” file to make UROSTYM. • Check the digital board sure the correct COM port is selected. • installation.
  • Page 21 Check "Stop/Reset? botton not respond. assembly. • Ceck header assembly at J5 Error Message appears: Incorrect characters used in Names for currents, envelopes, and/or current/envelope/template name. templates cannot contain punctuation characters, such as commas or apostrophes… LABORIE UROSTYM Service Manual UST-SM01...
  • Page 22: Electromagnetic Compatability (Emc)

    7 ELECTROMAGNETIC COMPATABILITY (EMC) LABORIE UROSTYM Service Manual UST-SM01...
  • Page 23: Tables Of Guidance And Declaration Regarding Electromagnetic

    The use of accessories, transducers, and cables other than those specified by LABORIE may result in increased EMISSIONS or decreased IMMUNITY of the Urostym. The Urostym needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this EMC section.
  • Page 24: Iec60601-1-2:2007 Table 1 Requirements

    Emissions IEC 61000-3-3 interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the equipment or shielding the location. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 25: Iec 60601-1-2:2007 Table 2 Requirements

    <5% UT (>95% a battery. dip in UT) for 5 dip in UT) for 5 Power Frequency (50 Hz) Magnetic Power frequency magnetic Field 3A/m 3A/m fields of a typical location in a IEC 61000-4-8 typical environment. LABORIE UROSTYM Service Manual UST-SM01...
  • Page 26: Iec 60601-1-2:2007 Table 4 Requirements

    Interference may occur in the vicinity of known RF transmitting devices and equipment marked with the following symbol: NOTE 1: At 80 MHz, the higher frequency range applies LABORIE UROSTYM Service Manual UST-SM01...
  • Page 27: Iec 60601-1-2:2007 Table 6 Requirments

    Over the frequency range 150kH to 80MHz, field strengths should be less than 3 V/m. 7.1.4 IEC 60601-1-2:2007 Table 6 Requirments: LABORIE UROSTYM Service Manual UST-SM01...
  • Page 28: Symbols Glossary

    (01) Global Trade Item imports the medical device. Number NOTE: Sterility symbols are applicable to consumbale devices only. Refer to consumable device instructions for use for (11) Date of Manufacture complete symbol and instructional overview. (21) Serial Number LABORIE UROSTYM Service Manual UST-SM01...
  • Page 29 9 SCHEMATIC/PCB LABOUT This section includes the schematics for the analog and digital boards as well as the board layouts for the UROSTYM. This will guide only qualified Laborie trained technicians on how to service the unit. 1. Board Schematics...
  • Page 30: End User License Agreement

    10 END-USER LICENSE AGREEMENT LABORIE UROSTYM Service Manual UST-SM01...
  • Page 31 LABORIE UROSTYM Service Manual UST-SM01...