Urostym™ Service Manual LABORIE UROSTYM Service Manual UST-SM01-V13.00...
Page 2
Tel.: +31.53.4803700 Fax: +31.53.4803701 TRADEMARKS Urostym is a trademark of LABORIE MEDICAL TECHNOLOGIES CANADA ULC. Other names may be trademarks of their respective owners. COPYRIGHT All rights reserved. No part of this publication may be reproduced in any form whatsoever without the prior written permission of Laborie Medical Technologies Canada ULC.
Manometry 0-250 cmH2O (0- ±10% FS 184mmHq) Output Channels: CAUTION: The values listed in the table below are the maximum values that the UROSTYM unit will supply. Laborie Medical Technologies strongly recommends staying within the specified ranges. Channel Range Tolerance ±...
EN ISO 15223-1 EN61000-3-2 • • EN ISO 14155 • EN 61000-3-3 • EN ISO 10993-1 • EN 61000-4-2:2009 Please refer to the EC Declaration of Conformity for a list of additional product related standards. LABORIE UROSTYM Service Manual UST-SM01...
All internal components are clearly identified and labeled below. Reference Number Description Fan Power Cables Analog Power Cables RS-232 Communications Cable UROSTYM Enclosure Digital Board Reset Button Cables Power Entry Module Power Board Ground Stud Terminal Analog Board LED Board LABORIE UROSTYM Service Manual UST-SM01...
5. Ensure the laptop is secured to the Urostym® by attempting to slide the laptop along the Urostym®. If the laptop slides remove the laptop from the Urostym® and remove the re- closable fasteners applied.
Velcro tape from the printer and cart shelf, clean the areas with alcohol wipes and retry the procedure from step 2. LABORIE UROSTYM Service Manual UST-SM01...
If the replacement fuse immediately burns out, move the power switch to the ‘off’ position and remove the power cord from the unit. Contact your LABORIE Service representative for further instruction. LABORIE UROSTYM Service Manual UST-SM01...
4 CALIBRATION CALIBRATE UROCAP USING THE UROSTYM CHECKDB UTILITY If the UDS120 software is not installed on the computer, the Urocap transducer can be calibrated through the Urostym CheckDB utility. LABORIE UROSTYM Service Manual UST-SM01...
To verify your calibration, choose a Uroflow protocol, set zeroes, and pour using the same pouring procedure as the calibration. The unit should provide you with the same volume reading as during calibration. LABORIE UROSTYM Service Manual UST-SM01...
5 CLEANING AND PREVENTATIVE MAINTENANCE UROSTYM UNIT AND ACCESSORIES The Urostym unit is not immersible. It should be wiped down with a clean cloth dampened with acleaning solution as per your facility’s cleaning instructions. • • • UROCAP UROFLOWMETER The Urocap may becoe dirty due to urine contamination and cleaning will be required.
DATA BACKUP Check the space available on the computer's hard disk at least every 6 months. Back up data from the UROSTYM MDB file located in the Urostym folder on the computer (C:\Program Files\Laborie Medical Technologies\Urostym\) on external media like CDs, DVDs or USB flash drives.
6 TROUBLESHOOTING GUIDE In Stage (2) of the “Recommendation” section on the following pages, the UROSTYM unit should be opened. Laborie Medical Technologies strongly recommends that only qualified Laborie trained technicians service the unit at this stage due to the high voltage present inside the unit. The tool to be used in this stage is a multimeter.
Page 20
DB9 connector to the digital between the PC and the • board. Check the “UROSTYM.ini” file to make UROSTYM. • Check the digital board sure the correct COM port is selected. • installation.
Page 21
Check "Stop/Reset? botton not respond. assembly. • Ceck header assembly at J5 Error Message appears: Incorrect characters used in Names for currents, envelopes, and/or current/envelope/template name. templates cannot contain punctuation characters, such as commas or apostrophes… LABORIE UROSTYM Service Manual UST-SM01...
The use of accessories, transducers, and cables other than those specified by LABORIE may result in increased EMISSIONS or decreased IMMUNITY of the Urostym. The Urostym needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this EMC section.
Emissions IEC 61000-3-3 interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the equipment or shielding the location. LABORIE UROSTYM Service Manual UST-SM01...
<5% UT (>95% a battery. dip in UT) for 5 dip in UT) for 5 Power Frequency (50 Hz) Magnetic Power frequency magnetic Field 3A/m 3A/m fields of a typical location in a IEC 61000-4-8 typical environment. LABORIE UROSTYM Service Manual UST-SM01...
Interference may occur in the vicinity of known RF transmitting devices and equipment marked with the following symbol: NOTE 1: At 80 MHz, the higher frequency range applies LABORIE UROSTYM Service Manual UST-SM01...
Over the frequency range 150kH to 80MHz, field strengths should be less than 3 V/m. 7.1.4 IEC 60601-1-2:2007 Table 6 Requirments: LABORIE UROSTYM Service Manual UST-SM01...
(01) Global Trade Item imports the medical device. Number NOTE: Sterility symbols are applicable to consumbale devices only. Refer to consumable device instructions for use for (11) Date of Manufacture complete symbol and instructional overview. (21) Serial Number LABORIE UROSTYM Service Manual UST-SM01...
Page 29
9 SCHEMATIC/PCB LABOUT This section includes the schematics for the analog and digital boards as well as the board layouts for the UROSTYM. This will guide only qualified Laborie trained technicians on how to service the unit. 1. Board Schematics...