Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Use of Electrical Surgical Knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them. The interference
can activate the OSD menu of the display and as such disrupt the functionality of the display.
Power connection – Equipment with external 24 VDC power supply
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC (
) SELV power supply.
The medical approved DC (
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
The equipment should be installed near an easily accessible outlet.
The equipment is intended for continuous operation.
The compliance of this monitor with Medical Safety and EMC requirements has been evaluated using the
external (optional) medical power supply model 'Skynet BAR-A159'. If a different power supply will be
used, further investigation for safety and EMC requirements have to be performed at system level.
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd. Ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Power cords:
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada: "hospital grade" cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equip. or on a
tag on the power cord
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
Mains lead protection: Power cords should be routed so that they are not likely to be walked upon or
pinched by items placed upon or against them, paying particular attention to cords at plugs and
receptacles.
The power supply cord should be replaced by the designated operator only at all time.
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle.
Liquids and moisture
Never expose the monitor to liquids or moisture.
Never use the monitor near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry
tub or in a wet basement.
The equipment is IP21 compliant (IPx5 front side only) in vertical position. The power supply is not
approved for IP21 and must be mounted in a flat position for best resistance to fluids.
) power supply must be powered by the AC mains voltage.