5.1.1 Login
1.
Touch the Login Icon to enter login screen
Enter "0000" in Password window
2.
3.
Touch the Check Mark icon to save and exit
5.1.2 Pressure Calibration
Note: Pressure calibration must be performed when the unit is activated for
the first time.
1.
Using a pressure gauge, verify that the input pressure line is delivering
approximately 100 psi
Note: If input pressure is not 100 psi, calibration will result in a
mismatch between the display pressure and the actual pressure on output
port B
2.
Once you have verified input pressure, touch the Calibration icon
to enter calibration screen.
7
Need help?
Do you have a question about the TS566R and is the answer not in the manual?
Questions and answers