July 2021
2.6
Adjust display orientation (optional)
To improve field access to wiring or to better view the optional LCD display:
Procedure
1. Loosen the set screw until the level detector housing can rotate
smoothly.
Do not unscrew all the way. Rotating the housing, without this screw
in place, can damage the internal wiring.
2. First, rotate the housing clockwise to the desired location.
If the desired location cannot be achieved due to thread limit, rotate
the housing counterclockwise.
3. Re-tighten the set screw.
Note
Do not attempt to rotate the display beyond the thread limits.
Figure 2-5: Housing Rotation
H3/32 in.
Quick Start Guide
Torque 30 in-lb (3 Nm)
Quick Start Guide
11