Control Accuracy; Measured Parameter Accuracy; Disposal - Respironics BiPAP AVAPS User Manual

Table of Contents

Advertisement

11.5 C
a
ontrol
CCuraCy
Parameter
IPAP
EPAP
CPAP
Breath Rate
Timed Inspiration
Ramp Duration
Rise Time
*
Limited to 25 cm H
**
Dynamic pressure accuracy is ± 5 cm H
and varying flow conditions. Static pressure accuracy is ± 2 cm H
with a Whisper Swivel II and no patient flow.
*** The range of values correspond to tenths of seconds (e.g., a setting of 4 indicates a Rise Time of 0.4 seconds).
**** Measured at the patient end of circuit with a Whisper Swivel II exhalation device and no patient flow.
11.6 m
P
easured
arameter
Parameter
Respiratory Rate
Exhaled Tidal Volume
Exhaled Minute Ventilation
Leak Rate
11.7 d
isPosal
Dispose of this device in accordance with local regulations.
User Manual
Range
4 to 30 cm
H
O*
2
4 to 25 cm
H
O*
2
4 to 20 cm
H
O
2
0 to 30 BPM
0.5 to 3.0 seconds
0 to 45 minutes
1 to 6 ***
O when using the Bi-Flex feature in S mode.
2
O measured at the patient end of the circuit with a Whisper Swivel II
2
a
CCuraCy
Accuracy
Greater of ±1 BPM or ±10% of reading when measured over a four
minute period
± (25 + 0.15 of reading) ml
± (1 + 0.15 of reading) L/min
± (5 + 0.15 of reading) L/min
Accuracy
± 5
cm H
O**
2
± 5
cm H
O**
2
± 5
cm H
O**
2
Greater of ± 1 BPM or
± 10% of the setting (when measured over a 4 minute
period)
± (0.1 + 10% of the setting) seconds
± 10% of the setting
±
25%****
O measured at the patient end of the circuit
2