Functions
5
Functions
5.1
Balancing system
With a balanced surgical microscope ARveo you can move the optics
carrier in any position without tilting or falling down.
After balancing all movements during operation only need a minor
force.
5.1.1
Balancing the optics carrier
On the optics carrier Leica M530 two directions of movement are
balanced: A and B.
5
4
The essential performance of the ARveo is:
Providing sufficient light to the situs and to guarantee mechanical
locking of the optics carrier in any possible positioning.
12
5.1.2
On the arm system the direction of movement C is balanced.
5.1.3
The parallelogramm balances the up/down movement (direction D).
1
If the direction D cannot be balanced, a weight disk has to be added
or removed, see section 7.6.4.
ARveo / Ref. 10 747 384 / Version 04
Balancing the arm system
Balancing the parallelogram
D
1
D