Selecting The Assistant; Stand Settings - Leica ARveo User Manual

Table of Contents

Advertisement

7.5

Selecting the assistant

7.5.1
Leica M530 with ULT530
Using knob (1) switch the light from the back assistant to the
X
side assistants.
1
ARveo / Ref. 10 747 384 / Version 04
7.6

Stand settings

7.6.1
Automatic balancing of the ARveo
WARNING
Danger of injury due to movement of the microscope during
the balancing process.
Do not sit or stand immediately next to the microscope
X
during the balancing process.
WARNING
Risk of injury due to downward movement of the surgical
microscope.
Complete all preparations and adjustments to the stand
X
before the operation.
Never change the accessories or attempt to rebalance
X
the microscope while it is above the field of operation.
Balance the ARveo after re-equipping it.
X
Do not release the brakes when the instrument is in an
X
unbalanced state.
Before re-equipping during the operation, first swing
X
the microscope away from the operating field.
Never carry out the intraoperative AC/BC balancing
X
above the patient.
WARNING
Danger of injury to the eyes due to possibly hazardous
optical infrared and UV radiation.
Do not look at the operating lamp.
X
Minimize exposure to eyes or skin.
X
Use appropriate shielding.
X
Switch on the microscope, see section 8.1.
X
Make sure that all accessories needed are installed and that
X
they are in the permitted weight range (see "Specifications" on
page 63).
Align the accessories in working position.
X
Preparation before surgery
25