MIDIMATE 602 & 622 Diagnostic Audiometers

7 Programming a User Test

7.1

Accessing user test programming mode

To access User Test programming, do as follows:
1. Switch off the MIDIMATE 602/622.
2. Press Softkey 3 while you switch on the power until the User Progr. Pa-
rameters menu is displayed.
3. Press Softkey 3, Progr. to select the Enable User Programming menu.
4. Press Softkey 1, Yes, or Softkey 2, No, as required.
5. Press Menu to quit the User Progr. Parameters menu and execute the
normal start-up procedure ending in User Test 1.
7.2
Programming a user test
A user test will store choices such as Output and Stimulus Side, the test fre-
quency and the hearing and masking levels, and also options such as HLL and PTS.
To program or change a user test, do as follows:
1. Activate user test programming as described in §7.1.
2. Press Menu once or twice to quit User Test 1 and display the Select Function
menu or select an existing User Test to be changed.
3. Set up or change the selected User Test as for manual setup by menu (§3.2).
4. To check that user test programming actually is activated, or to save the current
user test, press Softkey 3 to display the Options Menu. This menu should show
Progr. above Softkey 3.
5. Wait 5 sec or press Menu to return to the test mode, or press Softkey 3, Progr.,
to display the Programming User Test! menu.
6. Press Softkey 5, Quit, to change or check the test setup, or Softkey 1, Con-
tinue, to assign a number to the User Test.
7. Press Softkey 1, 2, 3, 4 or 5 to save the user test as Test1, Test2, Test3, Test4
or Test5 respectively, and exit from user test programming to test mode with
the setup of the user test you have just stored.