Testing the hearing instrument with OTOsuite HIT 3.9.1 How to perform a standard test 3.9.2 How to test the directional microphone HIT FreeStyle testing Navigating in HIT FreeStyle Cleaning, Repair and Calibration Cleaning 5.1.1 Cleaning the device Otometrics - AURICAL HIT...
Page 4
5.3.1 How often is calibration necessary? 5.3.2 Environment Report Procedure 5.3.3 Calibrating the coupler microphone The battery simulator Troubleshooting Standards and Safety AURICAL HIT Warning notes The OTOsuite HIT Module Manufacturer 7.4.1 Responsibility of the manufacturer Technical specifications Index Otometrics - AURICAL HIT...
It is assumed that the user has a basic knowledge of how to compare the results of the hearing instrument tests with the specifications from the hearing instrument manufacturer and to detect typical malfunctions of the hearing instrument. Otometrics - AURICAL HIT...
Indicates that you should take special notice. • 1.2.1 Navigating this manual Menus, icons and functions to select are shown in bold type, as for instance in: Set options Tools Options • Click the icon on the toolbar or select > Otometrics - AURICAL HIT...
Calibrating the reference microphone ► 15 Calibration ► 34 Calibration procedures and general recommendations about calibration are described in The test chamber Using AURICAL HIT and positioning hearing instruments in the test chamber are described in Testing hearing instruments ► 15 Otometrics - AURICAL HIT...
The coupler assembly The coupler assembly consists of the following parts: A. Coupler adapter B. Coupler cavity C. Coupler microphone Coupler adapter The Accessory Box provides a range of adapters for easy positioning of different types of hearing instruments. Otometrics - AURICAL HIT...
Page 10
B. ITE, RIE, thin-tube testing - High coupler position The Accessory Box Connect the mini-jack cable from the Accessory Box to the mini-jack socket under AURICAL HIT, and insert the coupler microphone in the microphone socket in the Accessory Box. A. Wireless hearing instrument testing Otometrics - AURICAL HIT...
To close the lid, place your hand on the lid and press gently until the closing mechanism takes over the action. 2.3.12 Telecoils 4 induction loops are fitted in AURICAL HIT for testing telecoil functioning: BTE testing ITE testing Otometrics - AURICAL HIT...
The handle is designed for carrying AURICAL HIT. Caution If you carry AURICAL HIT by its handle, do not use your other hand to support it by the lid, as this may • cause the lid to open and squeeze your fingers. Otometrics - AURICAL HIT...
The Accessory Box serves as holder for the coupler micro- phone for testing hearing instruments outside The coupler assembly ► 9 AURICAL HIT. See Caution Do not place the Accessory Box inside • AURICAL HIT during transportation. Otometrics - AURICAL HIT...
Testing a hearing instrument involves the following main tasks: Calibrating the reference microphone Otometrics recommends that you calibrate the reference microphone daily or weekly. Set up the interval to suit your pur- Calibrating the reference microphone ► 15 poses. See...
• directional microphone test. Positioning the hearing instrument – Angle or turn the hearing instrument so that the front microphone (1) faces the main loudspeaker as it would when worn on a person’s ear. Otometrics - AURICAL HIT...
Page 17
As a rule of thumb, position the reference microphone as close as possible to the front microphone of the hearing instrument without actually touching it. Maximum permitted distances are: Vertically (Y axis) 8 mm (above) Sideways (X axis) ±12 mm Back-to-front (Z axis) ±3 mm Otometrics - AURICAL HIT...
1. Select a battery simulator and insert it in the hearing instrument. With AURICAL HIT you receive a set of color-coded battery simulators, which are used to power the hearing instru- ment. They are also used as probes for measuring the power consumption. Color code Size ANSI PR63 7012ZD Otometrics - AURICAL HIT...
3. Click the adapter with hearing instrument into place on the coupler and place the coupler in the low position in AURICAL HIT. Otometrics - AURICAL HIT...
6. If you place some acoustic putty on the receiver wire, this will shift its resonance frequency and prevent the wire from vibrating and creating feedback during testing. 7. Click the adapter with hearing instrument into place on the coupler and place the coupler in the high position in AURICAL HIT. Otometrics - AURICAL HIT...
In the Canal). General guidelines for positioning the hearing Positioning the hearing instrument for directional testing is described in instrument ► 16 1. Wrap a thin roll of acoustic putty around the receiver end of the hearing instrument. Otometrics - AURICAL HIT...
Page 23
4. Align the microphones along the loudspeaker axis. 5. Position the reference microphone as close as possible to the front microphone of the hearing instrument without touching. 6. Close the lid. You are ready to start testing. Otometrics - AURICAL HIT...
Traditional BTE hearing instru- ments ► 20 Thin-tube hearing instruments ► 21 ITE hearing instruments ► 22 . Leave the hearing instru- ment placed on the coupler assembly in AURICAL HIT. 2. Set the hearing instrument to FM+HI mode. Otometrics - AURICAL HIT...
Page 25
For FM transmitter with lapel microphone: – If the FM transmitter is equipped with a lapel microphone, run the cable into the test chamber and fix the microphone to the elevation plate. Otometrics - AURICAL HIT...
IEC standards. You can also customize the available options and save them as user tests or start-up settings. The procedure 1. Launch the fitting software for the hearing instrument so that you can control its parameters. 2. Launch OTOsuite and select in the Navigation panel. Otometrics - AURICAL HIT...
This is done automatically when you start a directional test. The signal used for this test is a flat spectrum Broad Band Noise, band-pass filtered between 750 Hz and 5 kHz, and presented at 70 dB SPL. Otometrics - AURICAL HIT...
Page 28
The numerical dir- Measurements ectionality result shown in the table indicates the average front/back difference in the measured frequency range. Measurements table also includes the adaptation interval used in seconds. Otometrics - AURICAL HIT...
> Customizing the test table 1. You can customize different measurement tables which you can save and reload as user tests. This also allows you to start a sequence of measurements. Otometrics - AURICAL HIT...
Page 30
You can click any existing measurement in the table to load its settings. You can then repeat the measurement by starting it at the bottom of the table. Deleting a measurement Delete Curve To delete the existing measurement, right-click the measurement and select . Note that for predefined sequences, the cleared test will remain in the table. Otometrics - AURICAL HIT...
Page 31
Gain axis Gain You can display the measurement gain by clicking the check box in the measurement table. The gain is calculated as the difference between the coupler microphone and the reference microphone in the test cham- ber. Otometrics - AURICAL HIT...
5.1.2 Cleaning accessories Adapters If needed, remove any acoustic putty residue, and use an alcohol based wipe to clean the adapter. Otometrics - AURICAL HIT...
7. Calibrate the coupler microphone using a 94 dB calibrator. See Calibrating the coupler microphone ► 36 5.2.3 Replacing floor lining Remove the old floor lining from the base area, and place a new lining in the same position. Otometrics - AURICAL HIT...
This will help meet standards, and will ensure continued reliability and accuracy. Reference microphone Otometrics recommends that you calibrate the reference microphone daily or weekly. Set up the interval to suit your pur- Calibrating the reference microphone ► 15 poses.
Page 35
(2), along the Z axis. You can ensure this by looking at the test chamber from the side when you adjust the reference microphone position for validation. 5. Close the lid. 6. Select Tools > AURICAL HIT Calibration > More > Environment Report Start 7. Click . The test lasts less than 1 minute. Otometrics - AURICAL HIT...
5 Cleaning, Repair and Calibration The report 5.3.3 Calibrating the coupler microphone Coupler microphone calibration must be performed by a certified service technician. What you need to calibrate • 94 dB sound level calibrator • ½ in microphone adapter Otometrics - AURICAL HIT...
Five different battery pills are available as optional accessories (see for details). Voltage range 0 to 3.5 V Current range 0 to 50 mA Ω Source impedance range 0 to 25.5 Battery supply Ring Sense voltage Shield Ground Otometrics - AURICAL HIT...
If noise enters the test chamber via the table or stand, try adding a heavy-duty mounting mat or replace the table or stand with a more suitable support. • If needed, place the PC and any other equipment on a separate table to prevent noise and vibrations from influencing sensitive tests. Otometrics - AURICAL HIT...
This requirement applies in the European Union. Do not dispose of these products as unsorted municipal waste. You can return your device and accessories to Otometrics, or to any Otometrics supplier. You can also con- tact your local authorities for advice on disposal.
The equipment must com- ply with IEC 60950. The OTOsuite HIT Module Complies with Medical Devices Directive 93/42/EEC and RoHS Directive (2011/65/EC). Otometrics - AURICAL HIT...
The electrical installation to which the equipment is connected complies with EN/IEC requirements. • The equipment is used in accordance with the instructions for use. The manufacturer reserves the right to disclaim all responsibility for the operating safety, reliability and performance of equipment serviced or repaired by other parties. Otometrics - AURICAL HIT...
8 Technical specifications Technical specifications Type identification AURICAL HIT is type 1082 from GN Otometrics A/S. Acoustic stimulus generation In closed test chamber Frequency response, re. 1 kHz, main loud- 125 to 200 Hz: ± 3.0 dB speaker (equalized) 200 to 2000 Hz: ± 1.5 dB 2000 to 5000 Hz: ±...
Need help?
Do you have a question about the Aurical Hit and is the answer not in the manual?
Questions and answers