Data Interface; Entering Patient Id; Printing Audiograms (Midimate 622 Only); Storing Test Results - otometrics Madsen Midimate 602 Quick Manual

Diagnostic audiometers
Table of Contents

Advertisement

MIDIMATE 602 & 622 Diagnostic Audiometers

8 Data Interface

8.1

Entering patient ID

You can enter an identification number for the patient when the MIDIMATE 602/622
is on-line to a computer database. Do as follows:
1. Toggle the Menu key on the front panel to display the Select Function
screen.
2. Press Softkey 5 to display the Patient screen.
3. If required, press Softkey 1 to reset the number displayed to zero.
4. To change the number displayed press Softkey 2. The number is automatically
increased by 1.
5. To change the existing number use the Frequency wheel to select the digit to be
changed and change the value with the Hearing Level wheel. An asterisk (*)
marks the selected digit. You can enter spaces and dashes (–) as well as the nu-
merals 0 to 9.
6. Press Menu to exit the patient ID function.
8.2

Printing audiograms (MIDIMATE 622 only)

You can store air and bone conduction and free-field threshold data for both ears and
print it out on a standard printer, a ZODIAC 901 or an external printer connected to
the ZODIAC 901 as described in §2.1.1. To set it up see the description in §8.3.
8.2.1

Storing test results

To store the results of a current test for subsequent printing out, do as follows:
1. In test mode press Softkey 3 to display the Options Menu.
2. Press Softkey 3, Store, briefly. If printout has not been set up as in §8.3, the
Store option will not be shown.
8.2.2

Printing an audiogram

To print out stored thresholds as an audiogram, do as follows:
1. In test mode select the Printer Menu by holding down Softkey 3 while Store
is shown above it.
2. Press Softkey 4, Plot.
25