123B/124B/125B
Users Manual

Relative Measurements

Zero Reference shows the present measurement result
with respect to the defined value. Use this feature to
monitor the measured value in relation to a known good
value.
To setup:
1.
Push  to open the Scope and Meter button bar.
2.
Push 1 to open the MEASURE A button bar.
3.
Push 2 to enable the relative measurement
(Zero On).
4.
Push 4 to close the MEASURE A menu and
button bar.
The relative measurement is now the main reading,
and the former main measurement is now the smaller
secondary reading position.
5.
Repeat Step 1 to Step 4 to disable the relative
measurement.