Specifications - WIKA air2guide A2G-85 Operating Instructions Manual

Ventilation duct sensor for co2 and temperature
Table of Contents

Advertisement

Available languages

Available languages

9. Specifications

9. Specifications
Ventilation duct sensor for CO
EN
Version
Measuring range
CO
2
Temperature
Accuracy
CO
2
Temperature
Supply voltage U
Power consumption
Electrical connection
Output signal
Material
Case
Cover
Probe sleeve
Mounting flange
Permissible temperatures
Operation
Ambient
Relative humidity
Ingress protection per
IEC/EN 60529
Weight
Mounting
24
Version without LC display
Version with LC display
400 ... 2,000 ppm
0 ... 50 °C [32 ... 122 °F]
±40 ppm +2 % of reading value
< 0.5 °C [0.9 °F]
AC 24 V or DC 24 V ±10 %
B
Max. 230 mA
Cable gland M16
Screw terminals
Max. 1.5 mm
DC 0 ... 10 V, load min. 1 kΩ
4 ... 20 mA, load R min. 20 Ω, max. 500 Ω
Modbus
Plastic (ABS)
Polycarbonate
Plastic (ABS)
LLPDP
0 ... 50 °C [32 ... 122 °F] (at sensor)
-20 ... +70 °C [-4 ... +158 °F]
0 ... 95 %, non-condensing
IP54
150 g
Via adjustable mounting flange
WIKA operating instructions model A2G-85
and temperature, model A2G-85
2
2
®

Advertisement

Chapters

Table of Contents
loading

Table of Contents