Controls, Connections and Setup Procedures
6.2
6.2.1
Chapter 1 / 22
Start-up Procedure
Mechanical Start-up
ADC Compact Plus must not be connected to a network during the start-up
process.
Mechanical Start-Up
Start-Up:
1. Switch on the digitizer.
2. Check the operation terminal during the initialization.
While booting for the first time, the message
<Installation not yet confirmed> appears.
3. Confirm with OK
4. After a successful start-up <ready> appears.
5. Check/adjust date and time in key operator mode.
6. Choose terminal language in the service menu
<Configure>
<User terminal language>
7. Store the modified machine-specific data.
Check:
8. Open the service menu
<Maintenance>
<Handling Cycle>
and check the function of cassette- and plate operation with the machine
open. Each size should run at least one time.
9. Erase all image plates.
End of Mechanical Start-Up (approx. 30 min.)