13.
Decontamination and Cleaning
The following guidelines are suggested by Drive
DeVilbiss Healthcare Ltd. as being suitable
infection control procedures. Further information
is available upon request.
Always disconnect the mattress
system and bed frame from
the main power supply prior to
cleaning.
The control unit is rated to
IP21 and provides protection
from condensation only, do not
immerse or soak the control unit
– Risk of electric shock.

13.1 Control Unit

Use a cloth dampened with a mild detergent
and warm water (40˚C) solution.
This may be followed either by wiping with
a sodium hypochlorite solution to a dilution
of 1,000ppm/0.1% or by using an alcoholic
wipe.
Wipe thoroughly with clean cold water and a
damp single use wipe. Remove any excess
sodium hypochlorite solution.
Dry off with a paper towel - Always ensure
the cleaned surfaces are allowed to fully dry
before putting back into use.
Do not use phenolic, biological
or phenol based cleaning
solutions.
Do not use any abrasive
compounds or cleaning pads.
Do not submerge the control
unit into water or any other
liquids.
18
|
Drive DeVilbiss Healthcare Ltd
|
Hybrid-Power
|
Instructions for use

13.2 Mattress

Personal Protective Equipment
should be worn during the
entire cleaning process.
Regular cleaning and
disinfection of the mattress
system will help to prevent the
risk of infection to the occupant
and/or carer.
Prior to transferring the mattress
system to another user, ensure
it has been cleaned and
disinfected using the method as
detailed to help prevent the risk
of cross infection.
Inspection:
Before attempting to clean the mattress, the
top cover is to be checked for physical signs of
damage that may lead to strikethrough (ingress
of fluid through the cover). This is achieved by
unzipping the top cover and looking for signs of
staining to the white underside.
Replace the cover if strike-
through is evident – Risk of
cross infection.
Replace the cover if there are
any signs of damage (tearing,
punctures, abrasion, damaged
seams etc.)
General Cleaning:
Wipe down with a clean cloth moistened
with a mild detergent and diluted in warm
water (40˚C).
Rinse with cold clean water and a clean
cloth.
Dry off with paper towels - Always ensure
the cleaned surfaces are allowed to fully dry
before putting back into use.
|
2021-01 rev7