Digital flow switch for large flow
Table of Contents

Advertisement

Selection of the display on the sub screen
W
The display on the sub screen in measuring mode can be set.
Sub
screen
Switch output/Communication mode display
Displays the current mode
(Only for the IO-Link compatible products)
Analog output free range function
W
This function allows a flow that generates an output of 5 V (or 10 V when 0 to 10 V is selected) or 20 mA to be changed.
The value can be changed between 10% of the max. value of the rated flow and the max. value of the display range.
5
1
0
30
300
10% of the rated
flow range
Flow [L/min]
3000 L/min type
Error display function
W
When an error or abnormality arises, the location and contents are displayed.
Display
(Flashing)
(Flashing)
If the error cannot be solved after the instructions above are performed, please contact SMC for investigation.
Accumulated value display
Displays the accumulated value
Bottom value display
Displays the bottom value
Can be changed
3150
3000
100% of the rated
Max. value of
flow range
the display range
Error name
OUT over current error
A load current of 8 0 mA or more has been
applied to the switch output (OUT).
* Er2: PF3A8
-L series only
w
Instantaneous flow error
The flow rate, pressure, or temperature ex-
Pressure/Temperature error *
1
ceeds the upper limit of the setting range.
*1 PF3A8
-L series only
w
Pressure/Temperature error
The pressure or temperature exceeds the
lower limit of the setting range.
* PF3A8
-L series only
w
The accumulated flow has exceeded the accumulated
Accumulated flow error
flow range. (For accumulated increment)
The accumulated flow has reached the set accumulated
Accumulated flow error
flow value. (For accumulated decrement)
Outside of zero-clear range
During zero-clear operation, a pressure of 7 % F.S. or more has been
applied. (The mode is returned to measurement mode after 1 s.)
* PF3A8
-L series only
w
System error
An internal data error has occurred.
Version does not match
The IO-Link version does not match that of
the master.
* Only for the IO-Link compatible products
PF3A
Function Details
Set value display
Displays the set value
Line name display
Displays the line name
For analog voltage output of 0 to 10 V
10
0
30
300
10% of the rated
flow range
Description
Eliminate the cause of the over current by turning OFF
the power supply and then turning it ON again.
Decrease the flow rate, pressure, or temper-
ature.
Increase the pressure or temperature.
Reset the accumulated flow.
Retry the zero-clear operation without pres-
sure.
Turn the power OFF and then ON again.
Ensure that the master IO-Link version
matches the device version.
H(-L)
w
Series
Peak value display
Displays the peak value
OFF
Displays nothing
Can be changed
3000
100% of the rated
Max. value of
flow range
the display range
Flow [L/min]
3000 L/min type
Action
3150
38