Changing Chart Paper; Power Supply - Thermo Scientific TSX Series Installation And Operation Manual

Laboratory refrigerators
Table of Contents

Advertisement

10.2 Changing Chart

10.3 Power Supply

Thermo Scientific
Figure 16. Pressure Sensitive Chart Buttons
To change the chart paper, complete the following steps:
Paper
1. Locate the pressure sensitive buttons at the front, upper left of the
recorder panel.
2. Press and hold the change chart button (#3 in Figure 16) for
approximately one second. The pen will move off the scale.
3. Unscrew the center nut, remove the old chart paper, and install new
chart paper. Carefully align the day and time with the reference mark on
the recorder panel (a small groove on the left side of the panel, as shown
in Figure 15).
4. Replace the center nut and hand tighten. Press the change chart button
again (#3) to resume temperature recording.
Note Be careful not to pull up on the chart recorder arm while changing paper.
Pulling on the arm may damage the recorder.
CAUTION: Do not use sharp or pointed objects to depress the chart
buttons. This may cause permanent damage to the recorder.
The recorder normally uses AC power when the system is operating. If AC power
fails, the LED indicator on the recorder flashes to alert you to a power failure.
The recorder continues sensing cabinet temperature and the chart continues
turning for approximately 24 hours with back-up power provided by the
nine-volt battery.
The LED indicator glows continuously when main power is functioning and the
battery is charged.
When the battery is low, the LED flashes to indicate that the battery needs to be
changed.
TSX Laboratory Refrigerators
39