Startup; Initial Startup - Thermo Scientific TSX Series Installation And Operation Manual

Laboratory refrigerators
Table of Contents

Advertisement

8 Startup

8.1 Initial Startup

8.1.1 Additional
Chromatography Refrigerator
Thermo Scientific
To start up the refrigerator, complete the following steps:
1. Verify that the sensor probe(s) have been installed in the bottle(s) as
described in Section 7.7.
2. Connect the AC mains power cord to the mains power inlet connector.
3. Ensure the double pole circuit breaker switch located next to the power
inlet is in "ON" position (i.e. "I" position).
4. Insert the key in the switch and turn to the Power On position. The
display will show the actual cabinet temperature and the compressor and
evaporator should start within approximately
10 minutes.
5. Allow the unit to reach operating temperature before loading it with any
product. To stabilize the temperature profile, a 24-hour waiting period
is recommended.
6. If you desire to enable the alarms, turn the three position key switch one
turn further clockwise to the Alarm On position. To avoid nuisance
alarms, wait until the unit has cooled down to the desired operating
temperature (5°C default).
7. If you have a remote alarm, hook it up at this point (refer to
Section 7.9).
8. If desired, lock the cabinet door using the key. Place duplicate key copies
in a safe place.
All controls should now be fully operational, the alarm active (if enabled), and all
visual indicators active.
1. Connect the AC chromatography power cord to the chromatography
power inlet connector.
Startup
2. Toggle the Chromatography power inlet and Chromatography manual
reset switches to ON. If the internal outlet is GFCI protected, it may
need to be reset.
TSX Laboratory Refrigerators
27