TRIGger:ACQuire:LEVel:CURRent<NRf+>
Arguments
0 to MAX
Unit
A (amperes)
RSTValue:
0
Example
TRIG:SEQ2:LEV:CURR 5
TRIG:ACQ:LEV:CURR 2
Query syntax
TRIGger:SEQuence2:LEVel:CURRent?
TRIGger:ACQuire:LEVel:CURRent?
Returns
<NR3>
Related Command
TRIG:SEQ2:LEV:VOLT TRIG:SEQ2:HYST:CURR
TRIGger:SEQuence2:LEVel:DVM
TRIGger:ACQuire:LEVel:DVM
This command sets the trigger level for DVM measurements. A positive trigger
occurs when the input signal changes from a value less than the lower
hysteresis band limit to a value greater than the upper hysteresis band limit.
Similarly, a negative trigger occurs when the input signal changes from a value
greater than the upper hysteresis band limit to a value less than the lower
hysteresis band limit.
Command syntax
TRIGger:SEQuence2:LEVel:DVM<NRf+>
TRIGger:ACQuire:LEVel:DVM<NRf+>
Copyright © Itech Electronic Co., Ltd.
Trigger commands
78
Need help?
Do you have a question about the IT6400 Series and is the answer not in the manual?