Dos Modules; Connecting Dos Modules; Dispensing Liquids - Miele PG 8593 Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

Connecting DOS
modules
Dispensing liquids For adjusting dispensing concentration, see "Further settings/
Adding and dispensing process chemicals

DOS modules

If required, up to two additional, external dispensing modules (DOS
modules) for liquid process chemicals can be fitted retrospectively.
External DOS modules are fitted by Miele Service. Internal dispensing
systems cannot be retrospectively fitted.
DOS modules are supplied with their own installation instructions.
 Before fitting the DOS modules, compare the connection data
(voltage and frequency) on the data plates of the modules with that
on the data plate of your machine. If the data does not match, the
modules could sustain damage. If in any doubt, consult an
electrician.
 Power supply connection for DOS 3 neutralizing agent.
 Power supply for DOS 4.
 Connections for dispensing hoses.
Connect the module to the machine's power supply.
To connect the dispensing hose, release the hose clip on a free
connector and remove the safety cap.
Push the dispensing hose onto the connector and secure it with a
hose clip.
Unused connectors must be blanked off with safety caps to
prevent the leakage of wash fluid.
Dispensing systems."