Basic Operation; Switches & Knobs; Connections - PIE 830PM Manual

Panel mounted multifunction process calibrator
Table of Contents

Advertisement

PRACTICAL INSTRUMENT ELECTRONICS
MODEL 830PM
Process Calibrator
5
+ -
HI
Loop Diagnostics · Transmitter Supply
mA · V pH · TC · Ω · RTD · Hz · Pressure
HI
SET
LO
EZ-CHECK™
OVERLOAD
Hz SYNC
3
2
+
5
All mA Read V
EZ-CHECK™ SWITCH
q
SOURCE: Instantly output two preset settings by
moving the EZ-CHECK™ switch to the "LO" position
or "HI" position. For fast three point checks select
the "SET" position. The 830PM will remember
the last "SET" value, even with the power off.
These values can easily be changed to suit the
calibration requirements. The values stored in the HI
and LO positions are also used for Auto Stepping.
READ: Slide the switch to the SET position. The
830PM will display the current input value. Slide the
switch to HI and the highest value measured since
turn-on or reset will be displayed; slide the switch to
LO and the lowest value measured since turn-on or
reset will be displayed.
Test leads with shielded or non shielded banana
plugs may be plugged into the protected banana
jacks labeled 1 through 6.
PIE external pressure modules may be plugged
into the jack marked Pressure Module
Page 2

Basic Operation

SOURCE
6
20.000
mA
SOURCE
MAX
OFF
READ
MIN
READ
EZ-DIAL
Double Click
Menu
Push & Turn for
Fast Dialing
Push & Hold to
Store/Step
Pressure
piecal.com
Module
Source & Read mV
Hz
-2
+1
Source V
pH
1
4
RTD • Ohms
-
-
+
6
TC
Practical Instrument Electronics
Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
SOURCE/OFF/READ Switch
w
Select "SOURCE" to output mA, V, pH, T/C, Ω,
RTD or Hz.
Select "READ" to read mA, V, T/C, Ω, RTD, Hz or
pressure.
EZ-DIAL™ KNOB
e
SOURCE: Turn the knob to adjust the output level.
Turn clockwise to increase the output, counter
clockwise to decrease the output in one least
significant digit step at a time. Push down and turn
the EZ-DIAL knob for faster dialing.
Press and hold the knob for two seconds to store
desired EZ-Check™ HI/LO points in SOURCE
mode. Continue to press and hold the knob for two
more seconds to start the automatic ramping.
READ: Press and hold to transfer the current
temperature into the EZ-Check™ MIN/MAX
points. This clears the MIN/MAX readings
which will update as the input value changes.
Double Click the EZ-DIAL knob to change the
function of the calibrator and to select ranges,
units and other user settings.

Connections

Simulating or reading thermocouples requires
the use of thermocouple or extension grade
thermocouple wire terminated with a miniature
thermocouple plug.

Advertisement

Table of Contents
loading

Table of Contents