Invacare Perfecto 2 Series Service Manual page 80

Homefill compatible, sensored/non-sensored
Hide thumbs Also See for Perfecto 2 Series:
Table of Contents

Advertisement

SECTION 23—LEAK TEST
Check each end cap
for cracks and for
major leaks. You
should NOT use
leak detector on the
valve.
4 Way Valve
Use a 0-30 psi pressure gauge to monitor
system pressure swings, and also record the
amount of time it takes to achieve a shift (full
charge and full purge).
FIGURE 23.2 4 Way Valve Function Test
80
Perfecto
Part No 1148070
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Perfecto 2 irc5po2Perfecto 2 irc5p

Table of Contents