Compressor
To avoid personal injury or property damage, rotate the fan by hand prior to connecting
the unit to a power source. Check for suction at the air inlet by placing your finger over
the port as you rotate the fan. You should feel a slight suction with each rotation of the
fan. If you DO NOT feel a suction but you feel or hear a thump as you turn the fan, DO
NOT connect the unit to a power source. Review the assembly procedure for possible
error.
NOTE: For this procedure, refer to FIGURE 7.7 on page 42.
1. Reinstall compressor. Refer to Replacing Compressor Assembly on page 32.
2. Run unit and check for leaks around compressor hoses and compressor tubes. Refer
to Leak Test on page 77.
3. Take pressure readings to ensure proper system pressures. Refer to Checking Sieve
Bed Pressure on page 86.
FIELD SERVICE PARTS LIST FOR THOMAS 2660 SERIES COMPRESSOR
ITEM
PART
NO.
NO.
PER ASM
1
See Chart
2
624277
3
625776
4
626175
5
618114
6
See Chart
7
623143
8
625175
9
621591
638574
10
617177
11
617562
12
621485
13
621641
14
625094
15
623137
16
638281
17
638282
ITEM
PART NO.
1
607731
6
665167
Part No 1148070
WARNING
QTY.
QTY.
PER UNIT
2
1
2
1
2
1
2
1
2
1
2
8
2
1
1
2
2
4
2
4
1
2
2
4
2
1
1
2660CE37-989 COMPRESSOR
QTY.
2
1
SECTION 7—COMPRESSOR
DESCRIPTION
Connecting Rod, Eccentric & Bearing Assembly
Piston Cup
Screw - Piston Cup Retainer
Piston Cup Retainer
Cylinder Sleeve
Head
O-Ring - Head Gasket
Screw - Head
Valve Plate Assembly
Valve - Pressure Relief-Not Shown (115 V)
Valve Restraint
Valve Keeper Strip
Valve Flapper - Intake & Exhaust
Valve Plate
Screw - Valve Flapper
O-Ring Valve Plate
Fan - Gray
Fan - Black
DESCRIPTION
Connecting Rod, Eccentric & Bearing Assembly
Head
41