Invacare Perfecto 2 Series Service Manual page 9

Homefill compatible, sensored/non-sensored
Hide thumbs Also See for Perfecto 2 Series:
Table of Contents

Advertisement

Regulatory Listing:
Electrical:
Placement:
Tubing:
Relative Humidity:
Time of Operation:
Recommended Storage and Shipping Temper-
ature:
*NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30
minutes).
LABEL
SYMBOL
O
LABEL
SYMBOL
Part No 1148070
Double Insulated Product ETL certified complying with UL 1097
ETL certified to CSA C22.2 No. 68.
No extension cords.
No closer than twelve inches from any wall, furniture, draperies,
or similar surfaces.
7 ft cannula with a maximum 50 ft of Crush-Proof Tubing (DO
NOT pinch).
20 to 60%
Up to 24 hours per day.
-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.
IRC5PO2 INDICATORS
O
PURITY
2
SYSTEM OKAY
O
over 85%
2
2
O
Between 73% to 85% YELLOW Indicator light
2
SYSTEM FAILURE
O
Below 73%
2
IRC5P INDICATORS
STATUS
SYSTEM OKAY
SYSTEM FAILURE
TYPICAL PRODUCT PARAMETERS
INDICATOR LIGHTS
(LED)
GREEN Indicator Light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down.
INDICATOR LIGHTS
(LED)
GREEN Indicator Light
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician
9
Perfecto
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Perfecto 2 irc5po2Perfecto 2 irc5p

Table of Contents