Invacare Perfecto 2 Series Service Manual page 22

Homefill compatible, sensored/non-sensored
Hide thumbs Also See for Perfecto 2 Series:
Table of Contents

Advertisement

SECTION 4—TROUBLESHOOTING
SYMPTOM
Unit Operating,
Internal Status Indicators:
RED: Three Flashes
GREEN: One Flash
Control Panel Indicators:
RED: Off
YELLOW: Flashing
GREEN: On
Potential Obstruction Alert
does not activate on flows
less than 0.5 L/min.
Perfecto
2
PROBABLE CAUSE
Internal repairs required.
System leak.
Defective check valves.
22
SOLUTION
Replace SensO
circuit board. Refer to
2
Replacing P.C. Board on page 59.
Repair leak in product tank, regulator,
tubing, fittings, or flow meter.
Replace check valves. Refer to Replacing
Check Valves on page 47.