Low-Level Protection And Testing - Lauda Ecoline RE 204 Operating Instructions Manual

Low-temperature thermostats
Hide thumbs Also See for Ecoline RE 204:
Table of Contents

Advertisement

Starting up
6.5.2

Low-level protection and testing

$ $ $ $
15.05.01
YACE0060
− If necessary repeat several times in case that the pump does not purge
immediately.
− Testing at regular intervals by lowering the bath level. Place a hose on the pump
connector and pump some of the bath liquid into a suitable container.
− Step 1 - 2 must follow.
− Bath temperature during this test not below 0°C or higher than 50°C, otherwise
danger of burn injuries !
− If there is any irregularity when testing the safety devices, switch off the unit
immediately and pull out the mains plug !
− Have the unit checked by the LAUDA service or the local service organisation!
− Double signal tone sounds, if the bath liquid falls so
much that the heater is no longer covered with liquid
completely
1. The display shows LEVEL (low-level) and the fault
triangle is flashing
! heating is switched off on both poles,
! pump and refrigeration system are switched off by
the electronics.
2. Top up the bath
Section 5.2 and reset with the key.
33