Lauda Ecoline RE 204 Operating Instructions Manual page 24

Low-temperature thermostats
Hide thumbs Also See for Ecoline RE 204:
Table of Contents

Advertisement

Preparations
Hose connections
a) Elastomer tubing
Tubing type
EPDM-tubing,
uninsulated
EPDM-tubing,
uninsulated
EPDM-tubing
insulated
Silicone tubing,
uninsulated
Silicone tubing
insulated
Viton
− EPDM-tube not for Ultra 350 and mineral oils!
− Silicone oil causes pronounced swelling of Silicone rubber ! never use Silicone
− Protect tubing with hose clips against slipping off.
18
Int. dia.
Ø mm
9
12
12
ext. dia.
35mm approx.
11
11
ext. dia.
35mm approx.
11
oil with Silicone tubing!
Temperature
Application
range °C
10 to 120
for all bath
except for
Ultra 350 and
mineral oils
10 to 120
for all bath
except for
Ultra 350 and
mineral oils
-60 to 120
for all bath
except for
Ultra 350 and
mineral oils
-30 to 100
water/glycol
-60 to 100
water/glycol
-60 to 200
for all bath
Ref. No.
RKJ 111
liquids
RKJ 112
liquids
LZS 021
liquids
water,
RKJ 059
mixture
water,
LZS 007
mixture
RKJ 091
liquids

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents