Materials; Refrigeration System - Lauda Ecoline RE 204 Operating Instructions Manual

Low-temperature thermostats
Hide thumbs Also See for Ecoline RE 204:
Table of Contents

Advertisement

Unit description
independent of the control function
A low-level protection switches off the heating on both poles in order to prevent dry operation of the
heater. The pump is switched off electronically. The overtemperature switch-off point is adjusted with a
tool on a potentiometer and is always limited to 5 °C above the operating temperature range. A floating
contact "Combination fault" is available (
All settings and fault messages are stored in the memory in case of a supply failure or when the mains
switch is set to OFF.
The tubular heater is controlled from a modified PID controller through a triac circuit specially designed
to be unaffected by supply variations and interference.
3.4

Materials

All parts which come into contact with the bath liquid are made from high-grade materials appropriate to
the operating temperature. These are rust-free stainless steel, PPS plastics and fluoride rubber.
3.5

Refrigeration system

The refrigeration system consists essentially of a hermetically sealed compressor. Heat of
condensation and motor heat are dissipated by a fan-cooled finned condenser. Fresh air is drawn in at
the front of the unit, warmed air is discharged at the back and to the sides. The ventilation openings
must not be restricted in order to ensure proper air circulation.
At working temperatures below approx. 30 °C the refrigeration system operates continuously to remove
a certain amount of heat, with the heater acting in opposition to provide automatically controlled heating
power.
The compressors are fitted with a temperature monitor which responds both to the compressor
temperature and to the motor current. In addition the cooling system is protected against excessive
pressure by a pressure monitor. The refrigeration unit is switched on either automatically or manually
through the operating menu .(
Types RE 207, RE 210, RE 212, RE 220 are additionally provided with a cooling capacity adjustment
which saves energy in many operating situations, reduces the load on the refrigeration system and
produces a better temperature control. This feature, too, can be operated either automatically or
manually.
Cooling curves (
18.03.03
YACE0060
Section 6.5.4.).
Section 6.4.1.).
Section 9. Technical data)
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents