DAKOTA ULTRASONICS MMX-6 DL Operation Manual page 25

Multi-mode ultrasonic thickness gauge
Table of Contents

Advertisement

Dakota Ultrasonics
8) Press the Probe key. The display will flash 1OF2. Repeat steps 3
through 7 on the second calibration point. The MMX-6 DL will now
display the sound velocity value it has calculated based on the
thickness values that were entered in step 7. The MMX-6 DL is
now ready to perform measurements within this range.
The MMX-6 DL has the ability to display measurements in both the
English (IN) and Metric (MM) systems. The following steps outline the
procedure for changing units:
1) Press ON/OFF key to power up the MMX-6 DL.
2) Press the MODE key to activate features and settings.
3) Press the UP or Down arrow keys to scroll to the unIt symbol.
4) Press the SEND key to toggle the status of the units - IN / MM.
5) Press the MODE key once again to return to measurement mode.
22
Find Quality Products Online at:
MMX-6 DL Modes
Changing Units IN / MM
Units in/mm