Transport:
Battery performance:
Battery over current protection:
Light emitting diode:
Average light intensity:
Output peak wavelength range:
Unit handpiece dimensions:
Unit weight:
Complies with the following Directives/Standards:
93/42/EEC
2002/95/EC
EN 60601-1
EN 60601-1-2
DIN EN ISO 10650-2
DIN EN 62471
IEC 60825-1
Ambient temperature:
Between -10 °C and +50 °C (14 °F and 122 °F)
Relative humidity:
Between 30% and 80% (non-condensing)
Atmospheric pressure:
Between 500 hPa and 1060 hPa
• Low self discharge technology results in a long
operating life.
• Battery is pre-charged and ready to use upon
purchase.
• Time for battery recharge: Approximately 3 hours.
• Quick Recharging Mode: allows a minimum of
5 minutes operating time after 10 minutes re-
charging.
Resettable fuse
3 W LED
Approximately 1000 mW/cm
Between 460 nm and 490 nm
(intensity maximum at 475 nm)
Length: 240 mm
Width: 31 mm
Handpiece with battery pack: 90 grams
Charging Base with power connector: 200 grams
Council Directive 93/42/EEC of 14 June 1993 concerning medical
devices amended by directive 2007/47/EC, annex 1
Restriction of the use of hazardous substances in electrical and
electronic equipment
2007 – Medical Electrical Equipment (General requirements for ba-
sic safety and essential performance)
2007 – Medical electrical equipment – Part 1-2: General require-
ments for safety and essential performance – Collateral standard:
Electromagnetic compatibility – Requirements and tests
2007 – Dentistry – Powered polymerization activators – Part 2:
Light-emitting diode (LED) lamps
2009 – Photobiological Safety of Lamps and Lamp Systems
1993 – Safety of Laser Products