5. Thoroughly wipe device using a single use cloth soaked with a bactericidal, viru-
cidal, fungicidal alcohol-based instrument disinfectant solution approved according
to local regulations and use according to instrument disinfectant manufacturer's
instructions for use. Pay special attention to device seams and insertions.
Item can be reprocessed for up to 6,000 times.
4.3 SmartLite
FOCUS Shields
®
NOTICE
Wrong cleaning or disinfection method.
Damage to SmartLite
1. Do not clean or disinfect with highly aggressive agents (e.g. solutions
4.3.1 Cleaning and Disinfection
1. Discard examination gloves according to local regulations and disinfect hands.
2. Use a new clean pair of examination gloves.
3. If device is visually contaminated, clean device with an instrument detergent com-
patible to the used instrument disinfectant solution. After cleaning, wipe device with
a cloth soaked with water to remove instrument detergent and dry device with a
clean, lint-free, single use cloth.
4. Thoroughly wipe device using a single use cloth soaked with a bactericidal, viru-
cidal, fungicidal alcohol-based instrument disinfectant solution approved according
to local regulations and use according to instrument disinfectant manufacturer's
instructions for use. Pay special attention to device seams and insertions.
SmartLite
FOCUS Shields can be reprocessed for up to 100 times.
®
4.3.2 Sterilization
The SmartLite
®
1. Sterilize SmartLite
273 °F/29 psi; sterilization time 3 minutes 30 seconds.
Items can be reprocessed for up to 60 times.
6
®
based on orange oil or acetone).
FOCUS Shields can also be sterilized.
FOCUS Shields by steam autoclave at 134  °C/2  bar =
®
FOCUS Shields.