Appendix
10
Appendix
10.1 Technical data
Turbidity and solids concentration measurement with sensor CUS 31, CUS 41
Display and measuring range
Sensor CUS 31 . . . . . . . . . . . . 0 ... 9999 FNU, 0 ... 3000 ppm, 0 ... 3 g/l, 0 ... 200%
Sensor CUS 41 . . . . . . . . . . . 0 ... 9999 FNU, 0 ... 9999 ppm, 0 ... 300 g/l, 0 ... 200%
Measured value resolution
Sensor CUS 31 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . auto range
Sensor CUS 41 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . auto range
1)
Measurement deviation
, display
1)
Reproducibility
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max. 0.2% of MR
TU signal input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . digital
TU signal output
Current range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 / 4 ... 20 mA
Measurement deviation
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max. 500 Ω
Output range . . . . . . . . . . . . . . . adjustable, ∆ 0.1 FNU, ∆ 1g/l, ∆ 0.1%, ∆ 10 ppm
Temperature measurement
Temperature sensor
. . . . . . . . . . . . . . . . . . . . . . . . . . . NTC, 30 kΩ at 25 °C
Display range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –5.0 ... +70 °C
Measured value resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1 °C
1)
Measurement deviation
, display
Temperature signal output (optional)
Current range . . . . . . . . . . . . . . . . . . . . . 0 / 4 ... 20 mA, galvanically separated
Measurement deviation
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max. 500 Ω
Output range . . . . . . . . . . . . . . . . . . . . . .
Limit contactor
Limit adjustment range . . . . . . . . 0 ... 9999 FNU, 0 ... 9999 ppm, 0 ... 300 g/l, 0 ... 200%
Switching hysteresis . . . 0.00 ... 99.99 FNU, 0.00 ... 99.99 ppm, 0.0 ... 99.9 g/l, 0.0 ...99.9%
Pickup / dropout delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 ... 2000 s
Contact type (switchable) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . MIN / MAX
1)
acc. to DIN IEC 746 part 1, for nominal operating conditions
44
. . . . . . . . . . . . . . . . . . . . . . max. 0.5% of MR
. . . . . . . . . . . . . . . . . . . . . . . . . max. 0.75% of MR
. . . . . . . . . . . . . . . . . . . . . . . . . 1.0% of MR
1)
. . . . . . . . . . . . . . . . . . . . . . . . . max. 1.25% of URV
liquisys CUM 252
adjustable, ∆ 10 ... ∆ 100% of URV