Endress+Hauser Liquisys M CUM 223 Operating Instructions Manual
Endress+Hauser Liquisys M CUM 223 Operating Instructions Manual

Endress+Hauser Liquisys M CUM 223 Operating Instructions Manual

Transmitter for turbidity and solids content
Hide thumbs Also See for Liquisys M CUM 223:

Advertisement

Quick Links

BA 200C/07/en/05.99
No. 51500274
Software version 2.0 or later
Liquisys M
CUM 223 / 253
Transmitter for
Turbidity and
Solids Content
Operating Instructions
Endress Hauser
Nothing beats know-how

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Liquisys M CUM 223 and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Liquisys M CUM 223

  • Page 1 Liquisys M BA 200C/07/en/05.99 No. 51500274 Software version 2.0 or later CUM 223 / 253 Transmitter for Turbidity and Solids Content Operating Instructions Endress Hauser Nothing beats know-how...
  • Page 2 Need information on the instrument? Please read the following chapters: General information Safety You wish to install the instrument. The required steps are described in this chapter: Installation You wish to operate or reconfigure the instrument. The operating concept is explained in these chapters: Operation Instrument configuration Interfaces...
  • Page 3: Table Of Contents

    Liquisys M CUM 223 / 253 UM253E00.CHP Table of contents General information ........... 2 Symbols used .
  • Page 4: General Information

    General information Liquisys M CUM 223 / 253 General information Symbols used Warning: Note: This symbol alerts to hazards which This symbol indicates important may cause serious injuries as well as items of information. damage to equipment if ignored. Ignoring this information may result in malfunction.
  • Page 5: Product Structure

    Liquisys M CUM 223 / 253 General information UM253E01.CHP Product structure You can identify the instrument variant by the order code on the nameplate. ENDRESS+HAUSER ENDRESS+HAUSER LIQUISYS-M turbidity / Trübung LIQUISYS-M turbidity / Trübung order code / Best.Nr. CUM 253-TS0115 order code / Best.Nr.
  • Page 6: Safety

    Safety Liquisys M CUM 223 / 253 Safety Intended application The CUM 223/253 transmitter is a field-tested • Chemical industry and reliable measuring transmitter used to • Pharmaceutical industry determine the turbidity and solids • Drinking water treatment concentration of liquid media.
  • Page 7: Monitoring And Safety Features

    Liquisys M CUM 223 / 253 Safety UM253E02.CHP Monitoring and safety features Safety features Monitoring features The transmitter is protected against external In the event of a system error or power failure, influences and damage by the following an alarm condition is signalled via a fail-safe design measures: fault signalling contact.
  • Page 8: Installation

    5). (see chpt. 3.3) Measuring system The complete measuring system comprises: Optional: • The Liquisys M CUM 223 or CUM 253 • Extension cable CYK 8 transmitter • Junction box VBM or RM • A sensor with an integrated temperature sensor and a permanently attached cable •...
  • Page 9: Dimensions

    Liquisys M CUM 223 / 253 Installation UM253E03.CHP Dimensions ENDRESS+HAUSER LIQUISYS GEHAEUS1.CDR REL1 ALARM REL2 REL1 REL2 Dimensions of Liquisys M Fig. 3.3 CUM 253 FELD1.CDR Inside of housing of Liquisys M CUM 253: 1 Removable electronics box 2 Partition plate 3 Terminal blocks FELD2.CDR...
  • Page 10: Mounting

    Installation Liquisys M CUM 223 / 253 EINBAU1.CDR Liquisys M CUM 223, EINBAU2.CDR Fig. 3.4 panel-mounted version Mounting 3.3.1 Field instrument Several mounting options are available for the Weather protection cover CYY 101 can be Liquisys M in the field instrument version: used for outdoor installation in conjunction with all mounting variants.
  • Page 11 Liquisys M CUM 223 / 253 Installation UM253E03.CHP Post mounting kit Universal mounting post CYY 102 Mounting kit for installation of field housing on Square tube for mounting of measuring horizontal or vertical pipes (max. ∅ 60 mm); transmitters; material: 1.4301 (SS 304);...
  • Page 12 Installation Liquisys M CUM 223 / 253 3.3.2 Mounting examples Pipe mounting of Liquisys MONT10.CDR Fig. 3.7 M field instrument Liquisys M field instrument Left: Wall mounting Right: Mounting with universal post and weather MONT2.CDR Fig. 3.8 protection cover Endress+Hauser...
  • Page 13 Liquisys M CUM 223 / 253 Installation UM253E03.CHP 3.3.3 Panel-mounted instrument The instrument is attached using the supplied tensioning screws (see Fig. 3.9). The required overall installation depth is approx. 165 mm. +0.5 Attachment of panel-mounted instrument max. 45 approx.
  • Page 14: Electrical Connection

    Installation Liquisys M CUM 223 / 253 Electrical connection Connection diagram The connection diagram depicted in Fig. 3.10 shows the instrument connections for turbidity sensor CUS 31 or CUS 41. The connection of the various sensors is shown in more detail in Figs.
  • Page 15 Liquisys M CUM 223 / 253 Installation UM253E03.CHP Connections of field instrument For connection, the measuring cables are introduced through the cable glands on the field instrument and connected according to the diagrams in Figs. 3.10, 3.11 and 3.12. Connection compartment sticker for field instrument U-AN01.CDR...
  • Page 16: Sensor Installation And Measuring Cable Connection

    Installation Liquisys M CUM 223 / 253 Sensor installation and measuring cable connection Measuring cable connection The sensor is connected via a special Note: screened, multi-core cable which is completely terminated and permanently • Connectors, cable ends and termi- connected to the sensor.
  • Page 17 Liquisys M CUM 223 / 253 Installation UM253E03.CHP Connection examples CYK 8 Sensor Connection of turbidity sensors CUS 31 and U2X3AN.CDR Fig. 3.14 CUS 41 Endress+Hauser...
  • Page 18: Operation

    Operation Liquisys M CUM 223 / 253 Operation Operator interface Display of active Liquid crystal display contact and key for for display of measured relay switching in values and manual mode configuration data 1 8 5 R262 Setpoint LED indicator for alarm...
  • Page 19: Key Functions

    Liquisys M CUM 223 / 253 Operation UM253E04.CHP Key functions CAL key REL key When the CAL key is pressed, the instrument The REL key is used in manual mode to prompts for the calibration access code (fixed switch between the relays and the manual setting: 22 for calibration or any other number start cleaning function.
  • Page 20: Auto / Manual Mode Of Operation

    Operation Liquisys M CUM 223 / 253 Auto / manual mode of operation Auto mode Note: In this mode of operation, the relays • Enable the manual mode by are controlled by the transmitter. entering access code “22”. REL key •...
  • Page 21: Operating Concept

    Liquisys M CUM 223 / 253 Operation UM253E04.CHP Operating concept Operating modes Measuring mode: Standard mode of operation, display of current measured values Calibration mode: Setup mode: Execution of Access to all calibration routine configuration settings Code input: Following input of...
  • Page 22 Operation Liquisys M CUM 223 / 253 Menu structure The configuration and calibration functions Note: are arranged in a menu structure by function groups. • If a change is made but not confirmed by pressing the ENTER The function groups are selected in the setup key, the previous setting is mode with the PLUS and MINUS keys.
  • Page 23: Access Codes

    Liquisys M CUM 223 / 253 Operation UM253E04.CHP Access codes All instrument access codes are fixed, i.e. they cannot be modified. Three access codes are distinguished (cf. Fig. 4.4): • Any code: Read mode access, i.e. all • Code 22: Access to configuration menus...
  • Page 24: Instrument Configuration

    Instrument configuration Liquisys M CUM 223 / 253 Instrument configuration Setup mode After power-up (connection to power), the instrument performs a self-test and then ➜ SETUP 1 (A) see chpt. 5.2.1, p. 24 enters the measuring mode. ➜ SETUP 2 (B) see chpt.
  • Page 25 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Contact function, limit contactor 2 MAX contact without delay Current outputs 1 and 2 4 ... 20 mA Current output 1: meas. value for 0 FNU 4 mA signal current Current output 1: meas. value for...
  • Page 26: Start-Up

    Instrument configuration Liquisys M CUM 223 / 253 Start-up After power-up (connection to power supply established), the user must make the following selections in the specified function groups: • Function group SERVICE (S) • Function group SETUP 2 (B) S1: Select language and exit function Adjust all the parameters in this group;...
  • Page 27 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Selection or range Coding Field Display Info Factory setting The Liquisys M automatically CUS31 Display of sensor CUS31 detects which sensor is connected CUS41 connected. sensor Measured value damping causes averaging over the specified...
  • Page 28: Current Outputs

    Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Data set 1 can not be changed 1 -> 2 (factory setting). However, it can Copy data sets 1 -> 3 be used as a basic for a 2 ->...
  • Page 29 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Selection or range Coding Field Display Info Factory setting 4-20 Selection of 4–20 mA O211 O211 current range 0–20 mA Sel.Range 0.0 FNU 0/4 mA value; 0.0 ppm Enter the measured value enter 0.0 mg/l...
  • Page 30: Monitoring Functions

    Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Selection of table 1 ... number of table O233 O233 value pair value pairs Sel.Elem. asign 0.0 FNU 0.0 ppm 0.0 mg/l Enter x value...
  • Page 31 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP 5.4.1 Alarm Selection or range Coding Field Display Info Factory setting Function group Alarm function settings. ALARM ALARM Stead = steady Stead Selection of The contact type selected here contact contact type only applies to the alarm contact.
  • Page 32 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Automatic start of This field does not exist for some cleaning function? errors; see chpt. 7.1. CleanTrig If <—-R is selected, the display...
  • Page 33: Relay Contact Configuration

    Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Relay contact configuration The relay contacts described below can be • Limit contactor for meas. turbidity value: R2 (1) selected and configured as required (up to • Limit contactor for temperature: R2 (2) four times with four contacts depending on •...
  • Page 34 Instrument configuration Liquisys M CUM 223 / 253 P controller: Used for simple linear control PD controller: Used for processes that purposes with small system deviations. Where require quick response and where peaks are major changes are to be controlled, to be corrected.
  • Page 35 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Checking and fine tuning of parameter settings using a recorder Actual Actual value value too large too small Time Time Actual Actual value value too small too large Time Time Actual...
  • Page 36 Instrument configuration Liquisys M CUM 223 / 253 Control characteristic for direct and inverted control action Field R236 offers two control characteristics for selection which have the effects shown in the following diagram. 100 % Direct 100 % Inverted Control characteristic of a...
  • Page 37 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP 5.5.4 Chemoclean function Just like the timer function, Chemoclean can Thus, irregular cleaning with different repeat also be used to start a cleaning cycle. cycles is possible, and cleaning times with However, Chemoclean supports different post-rinse times can be individually defined.
  • Page 38 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Switch function of R211 R211 All settings are retained. R2 (1) off or on Function 9999 FNU 9999 ppm Never set switch-on point and...
  • Page 39 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Selection or range Coding Field Display Info Factory setting 100.0 °C Enter switch-on 100.0 °C Never set switch-on point and R222 R222 temperature –5.0 ... 100.0 °C switch-off point to the same value!
  • Page 40 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting 1.00 Enter control gain 1.00 R233 R233 See chpt. 5.5.2, p. 31. 0.01 ... 20.00 See chpt. 5.5.2, p. 31. Each hold resets the Enter integral I component to zero.
  • Page 41 Liquisys M CUM 223 / 253 Instrument configuration UM253E5A.CHP Selection or range Coding Field Display Info Factory setting LC PV = limit contactor TU (1) LC °C = Configure Cleaning is performed using only Timer limit contactor T (2) R2 (4)
  • Page 42 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Enter pre-rinse 20 s R253 R253 Water is used for rinsing. time 0 ... 999 s PreRinse Enter cleaning 10 s Cleaning agent and water are...
  • Page 43: Concentration Measurement

    Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP Concentration measurement The Liquisys M can convert turbidity values to concentration values (standard feature). [kg/l] Since not all possible units or the corresponding conversions can be made available, the user can create a custom conversion table.
  • Page 44: Service

    Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting 1 ... number of Selection of triplet Any triplet can be edited. triplets in K4 Sel.Elem. asign Enter turbidity 0.0 % value entire meas. range...
  • Page 45 Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP Selection or range Coding Field Display Info Factory setting The setting is retained even in Manual hold case of a power failure. Man. HOLD Enter hold dwell 10 s period 0 ... 999 s Cont.Time...
  • Page 46: E+H Service

    Instrument configuration Liquisys M CUM 223 / 253 E+H Service Selection or range Coding Field Display Info Factory setting Function group E+H SERVICE E+H SERV Contr = controller (1) Contr Selection of Trans = transmitter (2) module MainB = mainboard (3)
  • Page 47: Interfaces

    Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP Interfaces Selection or range Coding Field Display Info Factory setting Function group INTERFACE INTERFACE Address HART: 0 ... 15 Enter address For communication only. Address PROFIBUS: 1 ... 126 Factory settings are printed in bold face;...
  • Page 48: Calibration

    Instrument configuration Liquisys M CUM 223 / 253 5.10 Calibration Possibilities of calibration This function group is used to calibrate the Note: transmitter. Four types of calibration are possible: • If the calibration procedure is aborted by pressing the PLUS and •...
  • Page 49 Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP However, this procedure is not a mandatory • The calibration must be performed in prerequisite for calibration. You can also take ascending order of concentration. three different samples of the process •...
  • Page 50 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Function group CALIBRATION CALIBRAT 3-pt = three-point calibration (1) Corr = three-point correction (2) Edit = edit Only the “Data” function can be...
  • Page 51 Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP Selection or range Coding Field Display Info Factory setting Abortion Warning Warning Abortion E045 E045 o.k. E084 E084 Calibration status o.k. C114 C114 is displayed E xxx Status 20 % 200 %...
  • Page 52 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting Enter current value from 100.0 concentration of C131 C131 C111 first calibration entire meas. range Concentr1 solution Enter current value from 330.0 concentration of...
  • Page 53 Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP Selection or range Coding Field Display Info Factory setting If C142 = E xxx, then only no or new (exception: calibration Store calibration warning E 84). C143 C143 results? If new, return to C.
  • Page 54 Instrument configuration Liquisys M CUM 223 / 253 Selection or range Coding Field Display Info Factory setting 101.4 Calibration point 1 Deviation relative to standard C161 C161 reference value is displayed sensor (100 %). Concentr1 99.3 Calibration point 2 Deviation relative to standard...
  • Page 55: Offset

    Liquisys M CUM 223 / 253 Instrument configuration UM253E5B.CHP 5.11 Offset The settings in the OFFSET function group can be used to calibrate the measurement to Calibrated value a reference measurement. This requires a linear shift of all values measured, i.e. the...
  • Page 56: Interfaces

    Interfaces Liquisys M CUM 223 / 253 Interfaces Not available at time of printing. Endress+Hauser...
  • Page 57: Maintenance And Troubleshooting

    Liquisys M CUM 223 / 253 Maintenance and troubleshooting UM253E07.CHP Maintenance and troubleshooting Definition of terms Maintenance means that all measures which Troubleshooting means that the cause of the will guarantee the safety of operation and problem is determined and eliminated.
  • Page 58 Maintenance and troubleshooting Liquisys M CUM 223 / 253 Equipment, spare parts, Problem Possible cause Remedy information – Impermissible instrument Switch instrument off and back on EMC problem: check line routing if Fixed, incorrect operating state (no response to problem persists, check for...
  • Page 59 Liquisys M CUM 223 / 253 Maintenance and troubleshooting UM253E07.CHP Equipment, spare parts, Problem Possible cause Remedy information – No central HART module Verify by looking at nameplate: Upgrade to LSCH-H1 / -H2 HART = -xxx5xx and -xxx6xx – No or wrong DD (device description) –...
  • Page 60: Problem Elimination Based On Error Messages

    Maintenance and troubleshooting Liquisys M CUM 223 / 253 Problem elimination based on error messages Error Contact Error current Cleaning trigger Display Measures Fact. User Fact. User Fact. User EEPROM Switch instrument off and back on, return — —* E001...
  • Page 61 Liquisys M CUM 223 / 253 Maintenance and troubleshooting UM253E07.CHP Error Contact Error current Cleaning trigger Display Measures Fact. User Fact. User Fact. User Parameter range for Decrease current output spreading. — —* E080 current output 1 too small Parameter range for Decrease current output spreading.
  • Page 62: Diagnosis And Corrective Maintenance

    Diagnosis and corrective maintenance Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance Definition of terms Diagnosis refers to the identification of Diagnosis based on the error table below and instrument malfunctions and defects. depending on difficulty and measuring...
  • Page 63 Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance UM253E08.CHP Problem Possible cause Tests and / or remedial measures Equipment, spare parts, personnel Display shows – Instrument, or module not CUM 223: Reinstall module Refer to assembly drawings in...
  • Page 64: Corrective Maintenance Of Liquisys Cum 223

    Diagnosis and corrective maintenance Liquisys M CUM 223 / 253 Corrective maintenance of Liquisys CUM 223 8.4.1 Exploded view 223_EXO2.CDR Item 426a Terminal for signal inputs/outputs (external) Item 340 End frame Item 426b Item 400 Terminal for alarm relay Special...
  • Page 65 Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance UM253E08.CHP 8.4.2 Dismantling of CUM 223 • Consider potential effects on process • Loosen the special screw (item 400) by when removing the instrument from service! turning it counterclockwise. • First pull off the terminal block (item 424b) •...
  • Page 66: Corrective Maintenance Of Liquisys Cum 253

    Diagnosis and corrective maintenance Liquisys M CUM 223 / 253 Corrective maintenance of Liquisys CUM 253 8.5.1 Exploded view 253_EXO2.CDR Electronics box with relay module Fuse is accessible with LSR1-x (bottom) and power electronics box installed supply unit LSGA/LSGD (top)
  • Page 67 Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance UM253E08.CHP 8.5.2 Dismantling of CUM 253 • Open and remove the connection Make sure that the module catches do not compartment cover (item 420). come unlatched while you push the •...
  • Page 68: Spare Parts Orders

    Diagnosis and corrective maintenance Liquisys M CUM 223 / 253 8.5.5 Special case: replacement of central module Note: • Assemble the instrument with the new module as described in chapter 8.4.3 or 8.5.3. Replacement central module LSCx-x • Start up the instrument and test its basic is supplied from the factory with functions (e.g.
  • Page 69: Corrective Maintenance Of Measuring System

    Liquisys M CUM 223 / 253 Diagnosis and corrective maintenance UM253E08.CHP Corrective maintenance of measuring system 8.8.1 Turbidity measuring instrument The CUS 31 and CUS 41 sensors cannot be • Measure the sensor operating voltage: simulated since the entire measured value approx.
  • Page 70: Accessories

    Accessories Liquisys M CUM 223 / 253 Accessories approx. Dimensions of junction Fig. 9.1 box VBM VBM.CDR Connection accessories • Junction box VBM • Extension cable CYK 8 Junction box for extension of measuring Non-terminated measuring cable for cable connection between sensor and turbidity sensors CUS 31 and CUS 41 instrument.
  • Page 71: Technical Data

    Liquisys M CUM 223 / 253 Technical data °C Ω UM253E10.CHP Technical data General specifications Manufacturer Endress+Hauser Product designation Liquisys M CUM 223 Liquisys M CUM 253 Operating mode and system design Measuring principle A CUS 31 or CUS 41 sensor is connected to the Liquisys M CUM 223 / 253 transmitter via the digital sensor interface.
  • Page 72 °C Technical data Liquisys M CUM 223 / 253 Ω Output (continued) Temperature signal output (optional) Current range 0 / 4 ... 20 mA, galvanically separated Max. resolution 700 digits / mA max. 500 Ω Load adjustable, ∆ 10 ... ∆ 100 % of measuring range...
  • Page 73 Liquisys M CUM 223 / 253 Technical data °C Ω UM253E10.CHP Accuracy Turbidity measurement with CUS 31 Measured value resolution 0.001 FNU, 0.01 ppm, 0.1 g/l, 0.1 % Display deviation ±2 % of measured value (min. 0.02 FNU) Repeatability ±1 % of measured value (min. 0.01 FNU)
  • Page 74: Appendix

    Appendix Liquisys M CUM 223 / 253 Appendix Edit mode: Edit mode: code 22 code 22 Read mode: Read mode: any code any code Endress+Hauser...
  • Page 75 Liquisys M CUM 223 / 253 Appendix UM253E11.CHP Endress+Hauser...
  • Page 76 Appendix Liquisys M CUM 223 / 253 Endress+Hauser...
  • Page 77 Liquisys M CUM 223 / 253 Appendix UM253E11.CHP Endress+Hauser...
  • Page 78: Index

    Index Liquisys M CUM 223 / 253 Index Connection examples ....15 A ........24 Connections of field instrument .
  • Page 79 Liquisys M CUM 223 / 253 Index UM253E12.CHP General......2-3 O ....... . . 26 General safety instructions .
  • Page 80 Index Liquisys M CUM 223 / 253 S ........42 Technical data .
  • Page 81 ❑ Endress+Hauser Ges.m.b.H. Brazil Philippines Poland Wien ❑ Samson Endress+Hauser Ltda. ❑ Endress+Hauser Polska Sp. z o.o. Brenton Industries Inc. Tel. (01) 8 80 56-0, Fax (01) 8 80 56-35 Sao Paulo Makati Metro Manila Warszawy Tel. (0 11) 5 36 34 55, Fax (0 11) 5 36 30 67 Belarus Tel.

This manual is also suitable for:

Liquisys m cum 253

Table of Contents