Splice - UCL SWIFT SWIFT KR12 User Manual

Optical fiber arc fusion splicer
Hide thumbs Also See for SWIFT KR12:
Table of Contents

Advertisement

Q & A

SPLICE

When an error message appears on the screen.
HOW TO
·Refer to the [Error message list] for detailed information.
Irregular or higher splice loss.
HOW TO
·Clean V-Groove, V-Block, reflector and object lens by referring
to [Maintenance of splice quality]. Replace electrode
by referring to [Electrode replacement]. Refer to the
"High estimated loss" from [Error message list].
· If optical fiber warps or is bent, place the optical fiber's bent
direction to face the bottom. Splice loss varies depending on
cutting angle, discharge condition and cleanliness level of
optical fiber. Splice loss is still high or irregular even after
implementin the above recovery measures, contact to UCLSWIFT.
Annual maintenance is required to keep splice quality.
Monitor is turned off suddenly.
HOW TO
·Press any key and check [Monitor Power Save].
The power of the splicer is turned off suddenly.
HOW TO
·Turn on the splicer again and check [Splicer Power Save].
Unable to change arc discharge amount and time.
HOW TO
·You cannot change arc amount and time in SM, NZ, MM or Auto
mode. A proper arc discharge amount can be maintained in these
modes by performing [Arc Calibration]. The Arc amount and time will
set automatically to prevent from being changed in other mode.
How to set pause.
HOW TO
·Refer to [Option Menu].
Difference between estimated splice loss and
measured splice loss.
HOW TO
·The estimated splice loss is just a result from computation
so it has to be used for reference only.
��