Warm Up Timer; Cal Purge Timer; System Security - Macurco TracXP TXP-WTA Instruction Manual

Wireless sensor transmitter
Table of Contents

Advertisement

TXP-WTA_Instruction_Manual_(Rev1.3)

9.2.1 Warm Up Timer

The WARMUP timer has a default setting of 60 seconds but may be between 0 – 254 seconds. The primary purpose
of the warmup timer is to allow sensor stabilization after power up.

9.2.2 CAL Purge Timer

The CAL PURGE timer has a default setting of 60 seconds but may be set between 0 – 254 seconds. The primary
purpose is to allow sensor stabilization after a Span calibration (see Section 5.3). A purge interval is needed after the
span calibration, because up-scale readings will linger until the span gas exits the TXP-WTA Sensor Transmitter.
Alarms are inhibited during the CAL PURGE interval.

9.3 System Security

The SYSTEM SECURITY Menu in Figure 8 4 offers two levels of protection.
A LOW level allows CAL MODE sensor calibrations but requires the 4-digit Pass Code prior to altering menus.
HIGH level locks the entire menu database, plus CAL Mode, until the correct Pass Code is entered.
LOW and HIGH security levels always allow viewing of configuration menus, but they may not be modified.
Contact Name is a 12-character ASCII field available for displaying a phone number, or name, of personnel who
manage the Pass Code.
Lost Pass Codes may be recovered by entering the locked security menu and holding the UP key for 5 seconds. The
4-digit code appears near the bottom of the screen.
REV – 1.3
Figure 9-3 – Clock &Calendar / Delay Timer Menu
Figure 9-4 – System Security Menu
[34-2900-0207-4 ]