Operation; Patient Comfort Controls And Monitoring; Modes - Joerns Arise 1000 User & Service Manual

Lal mattress replacement
Table of Contents

Advertisement

Joerns® Support Surfaces
Arise® LAL Mattress Replacement
Arise® LAL Overlay, Arise® 1000, Arise® 1000EX
• If the patient needs the extended length functionality
of the bed (80" - 88), complete the following
procedure. Note: This can be accomplished with the
patient on the bed:
1. Adjust the Bari10A bed frame with the knee
section in the flat position.
2. Extend the bed frame (see bed manual).
3. Un-strap the mattress foot straps from the
Bari10A bed frame.
4. Un-snap the pocket that contains the
extension foot cells so they are free
to inflate.
5. Attach the extended length mattress foot
straps to the extended bed deck (Note: Do not
over tighten).
6. Inflate the extension foot cells by turning the
lower left knob on the mattress control panel
(Figure 2) up. After 2-5 minutes the cells will
inflate. Leave the knob in this position to
maintain inflation.
10
© 2019 Joerns Healthcare • 6110052 RevG • 19-2230
Figure 2

Operation

Warning: For important precautions, see page two.
Caution: The patient's head should be positioned in
the center of the top section of the mattress. When
using the mattress system always ensure that the
patient is positioned properly within the confines of the
bed. Do not let any extremities protrude over the side
or between the bed rails when the mattress is being
used.

Patient Comfort Controls and Monitoring

(Figure 3)
Power Switch
The power switch is used to turn the power on and off.
Standby Light
The Standby indicator will illuminate when the unit is
plugged into a power outlet. Press the Power button
to turn the unit on and inflate the mattress. Therapy is
the default mode at startup. When the Standby light is
on, it may also indicate that there has been a power
interruption and the therapy control unit is ready to be
turned back on. Press the Power button and reset the
preferred mode of therapy and comfort level.

Modes

Autofirm Mode
Autofirm mode provides maximum air inflation
designed to assist both patients and caregivers
during patient transfer and treatment. The unit will
automatically return to the mode it was in prior to
Autofirm (either therapy or pulsate) in approximately
12 minutes.
The Arise
has two low air loss therapy modes,
®
Therapy and Pulsate:
• Therapy: The unit starts in the Therapy mode,
which is the standard low air loss therapy.
• Pulsate: Pulsate will slightly decrease the pressure
in all cushions every 30 seconds then return to the
programmed comfort adjust level.