17) Troubleshooting

No measurement or unexpected value
Re-Used SmartPeg.
The SmartPegs are disposable and should only be used for one or
multiple measurement at one treatment session, for a single patient
only. Repeated re-use may result in false readings due to wear and
tear of the soft aluminium SmartPeg threads.
Wrong SmartPeg type selected for the Implant.
See SmartPeg reference list, osstell.com/smartpegguide
Bone or soft tissue in between SmartPeg and Implant.
Make sure to clean the Implant prosthetic connection before
attaching the SmartPeg.
Electromagnetic interference.
Remove the source of electromagnetic interference.
Instrument tip is held too far away from the SmartPeg.
Normally it is sufficient to hold the instrument tip 2-4 mm away from
the SmartPeg, but in some cases as close as 1 mm is necessary.
Instrument does not sense the SmartPeg, hence no measurement.
Perform a 2 cm circular movement of the instrument and then try to
measure again.
Unit is not charging when USB cable is connected
Wrong USB cable used.
Only use Osstell USB cable, type A-C, connected to a standard USB 2.0
or 3.0 (type A) port (Max 5,2 VDC).
14
3.7 VDC
Use only Osstell USB cable (USB type C / USB
type A) connected to standard USB 2.0 or 3.0
(type A) port (Max 5,2 VDC).
210 x 35 x 25 mm
272 x 140,2 x 74,60 mm
0,07 kg
0,75 kg
-40 ºC to +70 ºC
10% to 95%
500 hPa to 1060 hPa
+10 ºC to +35 ºC
30% to 75%
700 hPa to 1060 hPa
IP20
25087-05 US