The instrument uses infrared light emitting diode (LED 860±30nm) as the light source and 90° scattering method, which is compliant with ISO 7027 and DIN EN 27027 standards for the determination of turbidity for water solutions. TN400 allows you to measure the turbidity of an aqueous sample both in laboratory and in the field.
3.2 Meter and Accessories Calibration Solutions: ① 0.0, 20.0, 100, 400, 800 NTU Carrying case ② ⑨ TN400 Turbidimeter ③ ① Microfiber cloth ④ Screwdriver (to remove battery cover) ② ⑤ Silicone oil (10 ml) ③ ⑥ AA Alkaline battery×4 ⑦...
— Calibration mode ① — Standard values ② — Operation guide ③ Calibration Menu — Finish calibration indicator ④ 3.4 Keypad Diagram- 3 Functions ⚫ Power on/off ⚫ The meter will automatically shut off 10 mins after last key press ⚫...
icon displays to ensure measuring accuracy; when icon displays, the battery must be replaced, otherwise the instrument can't work properly. To extend battery life, user can setup auto- power off timer in Parameter Setup P6 according to usage requirements 3. Use the screwdriver to remove the four screws on the battery cover and open the battery cover. Insert 4 pieces of “AA”...
① The silicone oil applied should not be too much. After wiping with micro-fiber cloth, please wipe with filter paper or high-quality tissue paper to clean off. Excessive residual silicone oil on the vial surface will affect the measurement accuracy. ②...
4) After calibration is done, press to confirm. The instrument will return to the calibration menu (see Diagram on the right side). The OK sign indicates that 0 NTU has been calibrated and prompt to calibrate 20 NTU. 5) Place the 20.0 NTU calibration vial in the sample vial holder, press to start 20 NTU calibration.
Calibration point selection: The instrument has been calibrated at full range before leaving the factory. Users can select 2 or more points as needed to calibrate, as long as the target measurement range is in between of the two calibration points. In calibration setup mode, press select the calibration point.
(approximately 18ml), see Diagram-4, tighten the vial cap. 2) Before measuring, gently shake the sample vial to make sample solution even and wait for air bubbles to disappear, see Diagram-5, wait for 2~5 minutes until bubbles eliminate. 3) Wipe the sample vial surface with a microfiber cloth. Ensure that the vial is dry, clean and free from smudges.
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at the option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to the responsibility of APERA INSTRUMENTS, LLC for a period of TWO YEARS for the instrument from the delivery.
Need help?
Do you have a question about the TN400 and is the answer not in the manual?
Questions and answers