Elapsed Time Meter (Hour Meter) - Invacare Perfecto2 V User Manual

Oxygen concentrator
Hide thumbs Also See for Perfecto2 V:
Table of Contents

Advertisement

Invacare® Perfecto₂™ V Oxygen Concentrator
O
2
Symbol
Purity
SYSTEM
OKAY
O
A
2
O
over
2
85%
O
2
between
B
73% and
85%
O
2
O
un-
A
2
known
B
28
Indicator
Explanation
Light
(LED)
No action is
GREEN
needed. Normal
light
for operation.
Immediately call
supplier. You may
continue to use
the concentrator
YELLOW
unless instructed
light
otherwise by
(Solid)
your supplier.
Be certain that
backup oxygen is
nearby.
Immediately
GREEN
call supplier.
light with
Oxygen sensor
YELLOW
is malfunctioning.
light
You may continue
flashing
to use the
concentrator.
O
2
Symbol
Purity
SYSTEM
FAILURE
O
C
2
below
73%
D

6.3.8 Elapsed Time Meter (Hour Meter)

The hour meter displays the cumulative number of hours
the unit has operated. Refer to3.1 Component Identification,
page 14 for specific location.
Indicator
Explanation
Light
(LED)
Continuous
Audible Alarm
Sieve-GARD™
Compressor
Shutdown
RED light
SYSTEM FAILURE.
(Solid)
Total unit
shutdown.
Immediately
switch to backup
oxygen supply
and call supplier.
Call Supplier
1195056-B