Pressure Relief Mechanism
Operational at:
Change in maximum
recommended flow when
back pressure of 7 kPa is applied:
Filters:
Safety System:
Width:
Height:
Depth:
Weight:
Shipping Weight:
Operating Ambient Temperature
and Humidity:
Cabinet:
No AP/APG:
1195056-B
241 kPa ± 34.5 kPa (35 psi ± 5 psi)
0.7 L/min
Cabinet, Outlet HEPA and Compressor Inlet
Current overload or line surge shutdown
High temperature compressor shutdown
High Pressure Alarm w/ compressor shutdown
Low Pressure Alarm w/ compressor shutdown
Battery Free Power Loss Alarm
SensO
Oxygen System
2
Possible Obstruction Alert
38.1 cm ± 1 cm (15 in ± 3/8 in)
58.4 cm ± 1 cm (23 in ± 3/8 in)
30.5 cm ± 1 cm (12 in ± 3/8 in)
18.1 kg ± 1 kg (40 lb ± 2 lb)
20.4 kg ± 1 kg (45 lb ± 2 lb)
10°C to 35°C (50°F to 95°F) at up to 60% relative humidity
Impact Resistant flame-retardant plastic cabinet that conforms to UL 94 V-0
Not suitable for use in the presence of a flammable anaesthetic mixture.