Memory
2 x 60 memory spaces
Dimensions
L 97 mm x W 68 mm x H 22 mm
Weight
Approximately 108 g (without batteries)
Cuff size
135 to 230 mm
Permissible operating
+5 °C to +40 °C, 15-93 % relative hu-
conditions
midity (non-condensing)
Permissible storage
-25 °C to +70 °C, < 93 % relative humid-
conditions
ity, 700-1060 hPa ambient pressure
Power supply
2 x 1.5V
Battery life
For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
Data transfer
2402MHz – 2480MHz frequency band
Transmission power max. 0 dBm
The blood pressure monitor uses Blue-
tooth
Compatible with Bluetooth
phones/tablets
List of supported smartphones/tablets
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This device is in line with European Standard EN 60601-1-2
(In accordance with CISPR 11, IEC61000-4-2, IEC61000-
AAA batteries
®
low energy technology
®
4.0 smart-
4-3, IEC61000-4-8, EN 55022, EN 55024) and is subject to
particular precautions with regard to electromagnetic com-
patibility (EMC). Please note that portable and mobile HF
communication systems may interfere with this unit.
• This device corresponds to the EU Medical Devices Direc-
tive 93/42/EEC, the German Medical Devices Act (Medizin-
produktgesetz) and the standards EN 1060-1 (non-invasive
sphygmomanometers, Part 1: General requirements), EN
1060-3 (non-invasive sphygmomanometers, Part3: Sup-
plementary requirements for electro-mechanical blood
pressure measuring systems) and IEC 80601-2-30 (Medical
electrical equipment – Part 2-30: Particular requirements for
the basic safety and essential performance of automated
non-invasive sphygmomanometers).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life.
If using the device for commercial medical purposes, it must
be regularly tested for accuracy by appropriate means. Pre-
cise instructions for checking accuracy may be requested
from the service address.
• We hereby confirm that this product complies with the
European RED Directive 2014/53/EU. The CE Declaration
of Conformity for this product can be found under: www.
beurer.com/web/we-landingpages/de/cedeclarationofcon-
formity.php.
31

Advertisement

Table of Contents
loading

Table of Contents