Sampling
Minimum volume of
breath
Breath flow
Breath duration
Measurement accuracy
Standard deviation of repeatability using standard alcohol
<0.10 g/210 L
≥0.10 g/210 L
Up to 1.00 ‰
≥ 1.00 ‰
Wait time until next measurement can be taken following a pre-
vious measurement of:
0.05 g/210 L 20 s
0.10 g/210 L 40 s
0.20 g/210 L 80 s
>0.30 g/210 L 120 s
Test for accuracy
Dimensions
Weight
Power supply
Lithium-ion battery
AA batteries
Interface
LED risk group
Protection class
CE marking
Dräger Alcotest
Recommended: every 12 months
58 mm × 170 mm × 32 mm (W × H × D)
2.28 in × 6.68 in × 1.26 in (W × H × D)
approx. 245 g
approx. 8.64 oz
USB-C
2
IP 54
Electromagnetic compatibility
Low Voltage Directive
®
7000
≥1.2 L (default setting)
>10 L/min (default setting)
>1.6 s (default setting)
0.0042 g/210 L
1.7 % of the measured value
0.017‰ or
1.7 % of the measured value
Depending on which value is higher
0.50‰ 20 s
1.00‰ 40 s
2.00‰ 80 s
>3.00‰ 120 s
permanently installed in the device
optional
Disposal
|
en
27