Memory Operations

Datalog Operations

Datalog is a special application of store mode which enables the calibrator to
either automatically log displays at preset time intervals or to manually log a
display on operation of the STORE key. Logged data is written to a user specified
file.
To set up a datalog file, proceed as follows.
Select a task, other than BASIC. If using ADVANCED, set-up required Source
and Output parameters.
Use SETUP to select Datalog from the Store Mode Menu (See Page 36).

Auto Log (Timer)

Press STORE and set-up the datalog file parameters as shown below. Use CHANGE
VALUE (F1) followed by cursor keys to set field values. For Auto Log, set-up
TRIGGER field to PERIODIC.

Manual Logging

Enter the file details as shown above and select KEYPRESS for TRIGGER field.
Screen reverts to displayed parameters showing set-up file as shown below (e.g.),
Use a combination of STORE and LOG NOW (F1) to log events as follows.
K249 Issue No. 4