Pressure Ranges (Internal Transducers)
Safe working pressure
20 bar range
350 bar range
400 bar range
All other ranges
Accuracy
Combined non-linearity, hysteresis and repeatability
±70 mbar range
up to ±150 mbar
200 mbar to 20 bar (Calibrator):
35 bar to 700 bar (Indicator):
70 bar to 400 bar (Hydraulic)
Temperature Effects
±0.004% of reading/°C (averaged over -10° to -40°C w.r.t. 20°C)
Electrical Parameters
Voltage Inputs
Range:
Accuracy
Resolution
Current Inputs
Range:
Accuracy
Resolution
Current sink
Range:
Accuracy
Display
Size:
Reading
Environmental
Operating Temp.:
Storage Temp:
Calibration Temp:
Sealing
Sealed to IP54
Physical
Size:
Weight:
Specification
±30V
±0.05% rdg, ±0.004% F.S.
100µV max
±55mA
±0.05% rdg, ±0.004% F.S.
1µA max
24mA
±0.05% rdg, ±0.01% F.S.
60 x 60 mm LCD Graphics
±99999, update rate 2 readings/sec
-10°C to 50°C (Calibrated -10°C to 40°C)
-20°C to 60°C
21°C ±2°C
300 x 170 x 140 mm
3 kg
v
1.75 x full-scale
1.2 x full-scale
1.5 x full-scale
2 x full-scale
0.05% F.S.
0.05% span
0.025% F.S.
0.025% F.S.
0.025% F.S.