E4400-60229 Replacement Kit
Product Affected: . . . . . . . . . . . . . . . . . . . . . .
Serial Numbers: . . . . . . . . . . . . . .. . . . . . . . .
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Compatibilities: . . . . . . . . . . . . . . . . . . . . . . .
To Be Performed By: . . . . . . . . . . .
Estimated Installation Time: . . . . . . . . . . .
Estimated Performance Test Time: . . . . . .
Description
This ESG and ESG-D B Series RF Signal Generator Q501 Relocation Kit relocates Q501 to the
instruments chassis.
These instructions apply to all ESG models with a serial prefix <3934 for models
NOTE
E4400B, E4420B, E4421B, E4422B, E4430B/BU, E4431B/BU, E4432B/BU, and
E4433B/BU.
CAUTION
Failure to follow these instructions may result in a non-functional instrument.
To install this kit, the following major steps need to be performed:
1. Remove the existing A14Q501 from the CPU/Motherboard.
2. Install the 3 pin connector header into the CPU/Motherboard.
3. Relocate Q501 to the instrument's chassis.
4. Measure Q501's output voltage on the CPU/Motherboard.
After installing the hardware, perform the verification procedure as described in this installation note.
Installation Note E4400-90379
E4400B, E4420B, E4421B, E4430B/BU, E4431B/BU,
E4432B/BU, and E4433B/BU RF Signal Generators.
All
(X) Agilent Technologies Service Center
( ) Personnel Qualified by Agilent Technologies
( ) Agilent Technologies personnel on-site
1.0 hour
0.3 hour