Service Interface (X12); Shield Connector(S) For Physical Earth - IBA ibaPADU-8 Manual

Parallel analog digital converter for 8, 16 or 32 channels
Table of Contents

Advertisement

ibaPADU-8/ -16/ -32/ -32-R
8.2.5

Service interface (X12)

A 9-pin SUB-D port, found on the bottom side of the device is dedicated to load new
firmware on the device. For loading new firmware from a PC into the device you'll need
a standard V.24 interface cable.
Note
Please contact iba regarding loading new firmware. You will get the required files and
further information about the loading procedure.
Important note
In normal operation mode the V.24-cable must not be connected.
8.2.6

Shield connector(s) for physical earth

Proper connection of cable shielding to the device should be ensured. The shield con-
nector can be found on the bottom side of the device (or on the right side of each con-
nector on ibaPADU-16/-32/-32-R). It should be connected with any total shield that
serves for all cables. In addition, each individual sensor channel shield must be con-
nected to its respective input channel shield pin.
Note
Connect shield on one cable side only to avoid ground loops.
Note
Always ground the DIN rail on which the device is installed.