Returning For Service - DeFelsko PosiTector DPM Instruction Manual

Dew point meter
Hide thumbs Also See for PosiTector DPM:
Table of Contents

Advertisement

Find Quality Products Online at:

Returning for Service

Before returning the instrument for service...
1. Install new or newly recharged batteries in the proper
alignment as shown within battery compartment.
2. Examine the surface temperature sensor for dirt or damage.
3. If the humidity sensor is slow to respond or reading erratically,
recondition the sensor by removing the probe from the gage
body and wrapping it in a damp cloth overnight.
4. Perform a Hard Reset (pg. 4).
5. If issue is not resolved, Update (pg. 11) your PosiTector gage
body and re-attempt measurements.