Advanced Entry Calibration (continued)
Step
Action
Use the buttons to
increment the Gain
Number by one and then
J
press ENTER. Record the
TOF Value. Repeat this
process for each Gain
Number.
Gain Number
00
01
02
...
31
Step
Action
In your sample calibration
table, where both the
Label Value and TOF Value
are at least 20, subtract
the amounts and record
the result as a Difference
Value (see below).
K
Identify the largest
Difference Value and the
corresponding Gain
Number. This Gain
Number will be used to
resample the media.
Gain Number
00
01
02
...
15
16
17
18
19
...
31
84
Displayed Message
TRAN SENSOR GAIN
yyy
Sample Calibration Table
Label Value
TOF Value
252
248
250
245
248
234
...
09
Displayed Message
TRAN SENSOR GAIN
yyy
Sample Calibration Table
Label Value
TOF Value
252
248
250
245
248
234
...
188
184
179
174
170
...
132
Comment
These are TOF Values,
where 'yyy' represents
(0 - 31)
the current sensor
01
reading.
Difference Value
...
14
Comment
In this example, Gain
Number 18 is chosen
because, where both
(0 - 31)
values are at least
31
twenty, it has the highest
Difference Value.
Difference Value
4
5
14
...
...
63
125
51
133
38
141
25
149
19
N/A
...
...
14
N/A