Konica Minolta CM-23d Instruction Manual page 38

Table of Contents

Advertisement

2
Use [] or [] to move the cursor to "Calibration
(Including Zero Cal.)", and then press the
[Confirmation] key.
3
Set the instrument in the calibration stage to
measure the calibration stage's zero calibration
hole.
Calibration stage
4
Press the measurement button.
Zero calibration will be performed.
Notes
• Do not move the instrument until zero calibration is
complete.
When zero calibration is completed, a screen prompting
white calibration is displayed. Continue to step 3 on the
following page to perform white calibration.
35
Place the instrument
on the calibration stage
according to the marks.
Place the instrument
on the calibration stage
according to the marks.
Calibration stage
Place the instrument
on the calibration stage
according to the marks.
Calibration stage
Please set the
instrument on the zero
calibration box and
press the measurement
button.
CALIBRATION
Calibrating to zero